105
Connection
HBM
QuantumX
6.8.4
LED status display
System LED
CAN LEDs “BUS”
CAN LEDs “Channel”
System LED:
Green
Error
−
free operation
Yellow
System is not ready, boot procedure running
Flashing yellow
Download active, system is not ready
Red
Error, faulty synchronization
CAN LEDs (BUS):
Green flickering
No bus error and CAN activity
Constant green
No bus error and no CAN activity
Yellow flickering
Intermittent bus errors (Warning) and CAN activity
Constant yellow
Intermittent bus errors (Warning) and no CAN activity
Red on
Bus error, CAN interface in ”Bus
−
OFF” status
CAN LEDs (Channel):
Green flickering
No bus error and CAN activity
Constant green
No bus error and no CAN activity
Yellow flickering
Intermittent bus errors (Warning) and CAN activity
Constant yellow
Intermittent bus errors (Warning) and no CAN activity
Red on
Bus error, CAN interface in ”Bus
−
OFF” status
Ethernet LED:
Green on
Ethernet link status is ok
Flashing yellow
Ethernet data transmission ongoing
Содержание QuantumX MX1609
Страница 1: ...I3031 8 0 en Operating Manual ...
Страница 2: ...d ...
Страница 31: ...31 Housings HBM QuantumX Fig 4 2 Amplifier MX1609 P in IP65 housing ...
Страница 34: ...34 Housings HBM QuantumX a f 2 5 Fig 4 7 Mounting the CASEPROT ...
Страница 36: ...36 Housings HBM QuantumX Fig 4 10 Attaching the covers ...
Страница 38: ...38 Housings HBM QuantumX Fig 4 13 Close the lever Fig 4 14 Connected housings ...
Страница 99: ...99 Connection HBM QuantumX Connection of the thermocouple plug in the miniature design ...
Страница 144: ......
Страница 180: ...179 Support HBM QuantumX ...