Configuring the CX22‐W for automatic measurements
Settings in two areas must be implemented to configure the CX22‐W for
autonomous measuring:
1. You must implement the applicable settings in DAQ jobs and Options in
catman
®
AP so that catman
®
AP loads a DAQ project and launches the
DAQ job after startup.
2. You must ensure that catman
®
AP is started when the CX22‐W is
switched on.
Measurement start and stop can be implemented as usual immediately or via
a trigger. You can also use the START/STOP button of the CX22‐W, see Sec
tion 10.2 on Page 35.
Procedure
1. Configure the DAQ (data acquisition) project as described in Section 8.2.
Set up the DAQ jobs so that the data are saved with the
Automatically
on DAQ termination
setting (
Data storage
in the group
Settings
, para
meter
Data saving
).
2. Save the DAQ project.
3. Call up the catman
®
AP Options (
File
menu) and go to
Program start
(group
System
).
4. Select an
Existing DAQ project
for the
Start mode
and select the project
created above in the
DAQ project
line.
5. Activate the option
Execute first DAQ job automatically
.
6. Go to
DAQ channels
in the group
Channels and sensors
.
7. Activate the option
Deactivate defective channels and continue with
DAQ
(
If channel initialization fails
).
8. Exit the dialog by clicking on
OK
.
9. Exit catman
®
AP.
10. In the start screen, click on
Settings
.
11. In the
General
tab, activate the option
Launch DAQ software (cat
manAP) at startup
and click on
OK
.
12. Click on
Commit changes and reboot
to permanently save the new set
tings.
This project will then be automatically started when the CX22‐W is switched
on again. If the power supply is interrupted during measurement, the project
will be reloaded and measurement continued when the power supply is
restored. The new start of the module up to continuation of measurement will
require approx. 3 minutes, dependent on the number of connected devices
and channels.