14
P3MBP Blue
Line
A1710-3.3 en
HBM
6
Specifications (in accordance with DIN 16 086)
Type
P3MBP Blue
Line
Mechanical input quantities
Pressure type
Absolute pressure
Principle of measurement
Foil strain gage
Measuring range,
0 bar...
bar
5000
10000
15000
1)
Input characteristics
Accuracy class
0.2
0.4
0.5
Initial value
bar
0
Operating range
at reference temperature
%
120
Overload limit
at reference temperature
%
120
Test pressure
%
200
130
100
During dynamic loading
Permissible pressure
%
100
Permissible vibration amplitude
according to DIN 50 100
bar
3500
5000
6000
Dead volume
(with supplied HBM sealing)
mm
3
210
120
Control volume
mm
3
approx. 1
Output characteristics
Nominal (rated) sensitivity
mV/V
1
Sensitivity tolerance
%
t"
0.2
t"
0.4
t"
0.5
Effect of temperature on the zero signal
in the nominal (rated) range of the excitation
voltage per 10K, by reference to the nominal
(rated) sensitivity
in the nominal (rated) temperature range
in the operating temperature range
%
%
"
0.1
"
0.15
"
0.2
"
0.25
"
0.2
"
0.25
Effect of temperature on sensitivity
in the
nominal (rated) range of the excitation voltage
per 10K, by reference to the actual value
in the nominal (rated) temperature range
in the operating temperature range
%
%
"
0.1
"
0.3
"
0.2
"
0.4
"
0.2
"
0.4
Characteristic curve deviation
(start setting)
%
0.2
0.4
0.5
Repeatability
according to DIN 1319
%
t"
0.05
Fundamental resonance frequency
kHz
u
100
Input resistance
at reference temperature
Ω
350
"
5
Output resistance
at reference temperature
Ω
350
"
1.5
Isolation resistance at 100V AC
M
Ω
5000
Excitation voltage
Nominal (rated) excitation voltage
V
2.5
Operating range
V
0.5 ... 10
1)
available in 4rd quarter of 2003