![HBM MP07 Скачать руководство пользователя страница 41](http://html1.mh-extra.com/html/hbm/mp07/mp07_operating-manual_2111883041.webp)
41
PME-MP60/MP07
A0616-13.4 en
HBM
Group
Parameters
Meaning
CONDITIO
NING
>0< kNm
Enter zero value. Zeroing effects both, the gross value and
the net value.
>0< Set ?
Trigger zero balance; current measurement value
(physical unit) zero
>0< save
Each time there is a zeroing procedure the zero value is
adopted into the EEPROM (service life 100.000 cycles)
>T< kNm
1)
Input tare value. Taring has an effect on the net value.
>T< Set ?
Trigger taring; net value set to 0
>T<Save
Save tare value immediately after taring
Filter
0.05 Hz
1 Hz
20 Hz
500 Hz
0.1 Hz
2 Hz
50 Hz
Off
0.2 Hz
5 Hz 100 Hz
0.5 Hz 10 Hz 200 Hz
FiltChar
Step-function response
Best course over time
(Bessel)
Time
Time
Best frequency response
(Butterworth)
The diagram shows a linear amplitude
response that falls away steeply
above the cut-off frequency. An
overshoot of approx. 10 % occurs.
The diagram shows an step-function
response with a very small overshoot
(<1 %) or none at all. The amplitude
response drops off less steeply.
Amplitude response
1)
Depending on the unit selected