PME-MP60/MP07
A0616-13.4 en
HBM
The T10F is set by using sensitivity. To achieve greater accuracy, you must
use the integrated shunt to calibrate the T10F.
Calibration with integrated shunt:
The transducer must be load
−
free!
Group: TRANSD.-CALIBRAT
P1Meas.?
YES
(corresponds to zero point measurement)
P1 kHz
is displayed
P1 kNm
0 kNm
Shunt
ON
P2 Meas.?
Yes
P2 kHz
is displayed
P2 kNm
enter the calibration value in accordance with the
identification plate or the T10F test log
Rotation speed measurement
Measurement preparation:
"
Use the standard cable md-cab150.
"
Use the MD cable (cab149) to supply the transducer or use an external
supply voltage
"
Connect the transducer (see Connection
B
"
No
DIP switch position changes!
Defaults:
T10F at 360 increments/revolution
U = 6000 rpm
this corresponds to: 6000 U/min = 6000/60 U/sec = 100 U/sec
at 360 pulses/revolution:
f = 36 kHz
1)
In this example, the F2 signal (required for detection of direction of
rotation) is to be evaluated.
1)
choose the next highest frequency as the input range, here: 100 kHz