Profibus
ML77B
A00760_09_E00_00
HBM: public
27
11
DiagChn12
Error condition of channel 12
12
DiagChn13
Error condition of channel 13
13
DiagChn14
Error condition of channel 14
14
DiagChn15
Error condition of channel 15
15
DiagChn16
Error condition of channel 16
9.1.2
Measured values
Measured values can be transmitted in different formats such as IEEE float,
Siemens float (4 bytes), 16-bit fixed point number (16-bit integer as two's com
plement, decimal point position must be signalled to reading device) or 32-bit
fixed point number (32-bit integer as two's complement, decimal point position
must be signalled to reading device) (also see chapter
Parameterisation
). For
converting the values into fixed point format, the number of decimal places
must be specified in the parameterisation telegram.
9.1.3
Channel status for 1-channel device
Mode0
Bit
Name
Meaning
0
GW1
Status of limit switch 1
1
GW2
Status of limit switch 2
2
GW3
Status of limit switch 3
3
GW4
Status of limit switch 4
4
BRT OVLD
Gross value overload
5
NET OVLD
Net value overload
6
CAL_ERR
Calibration error or other error in channel
7
CHANGE
Changed scaling of measured values
8
PAR1
Active parameter set bit 1
9
PAR2
Active parameter set bit 2
10
PAR3
Active parameter set bit 3
11
PAR4
Active parameter set bit 4
12..14
Res
Reserved
15
MODE
0