Command set of the HBM interpreter
Amplifier settings
DMP41
A03474_04_E00_00
HBM: public
43
Effect:
Output of set filter
Response:
q1(y)
q1
Filter code number
1
fc1
2
fc2
Example:
AFS?(x)
1(y)
Filter fc1 is switched on
Amplifier Signal Filtering
Input of cut‐off frequency and filter characteristic
Syntax:
ASF p1,p2,p3(x)
Parameter:
p1
Filter code number
1
fc1
2
fc2
p2
Filter frequency
1...n
Code number for frequency value
(equivalent to the index in the frequency
table that can be output with the command
ASF?0).
p3
Filter characteristic
0
Bessel
1
Butterworth
Effect:
The low‐pass filter fc1 or fc2 is assigned a frequency value
and a filter characteristic.
(See frequency table for command ASF?).
Response:
ASF