![HBM DMP41 Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/hbm/dmp41/dmp41_interface-description_2111867037.webp)
Command set of the HBM interpreter
Communication
DMP41
A03474_04_E00_00
HBM: public
37
5.2.3
Error control, status register
Extended Status Query
Extended status query
Syntax:
XST?(x)
Parameter:
none
Effect:
The DMP41 responds with a bit‐coded decimal number.
Bit
Hex
Decimal
0
0001
1
1
0002
2
2
0004
4
3
0008
8
4
0010
16
5
0020
32
6
0040
64
7
0080
128
8
0100
256
9
0200
512
10
0400
1024
Bit0:
Transducer error
Bit1:
Calibration error
Bit2:
No calibration carried out
Bit3:
Timeout output stage
Bit4:
Amplifier overloaded
Bit5:
Sense lead (2-2') interrupted
Bit6:
Sense lead (3-3') interrupted
Bit7:
Calibration running
Bit8:
Background calibration running
Bit9:
Filter settling time
Bit10:
Background calibration error
Bit11:
No 0‐wire detected
Bit12:
No 1‐wire detected
Bit13:
Converters (internal) not synchronous
Bit14:
Initialization (new transducer)
Bit15:
Modulator overloaded
XST? outputs the sum of all relevant bits.
Response 258 means: 256 (calibration) and 2 (calibration error). This is a
typical response after a channel change and ongoing initial calibration.
XST?
Содержание DMP41
Страница 1: ...Interface description and command set English DMP41 Digital precicion measuring device...
Страница 10: ...Interface comparison 10 A03474_04_E00_00 HBM public DMP41...
Страница 14: ...Communication with the DMP Ethernet interface 14 A03474_04_E00_00 HBM public DMP41...
Страница 18: ...Communication with the DMP RS 232 interface 18 A03474_04_E00_00 HBM public DMP41...
Страница 22: ...Program creation Communication via the Ethernet interface Test program 22 A03474_04_E00_00 HBM public DMP41...
Страница 86: ...Functional command overview 86 A03474_04_E00_00 HBM public DMP41...
Страница 90: ...Index 90 A03474_04_E00_00 HBM public DMP41...
Страница 91: ...Index DMP41 A03474_04_E00_00 HBM public 91...