7.5 Complete the test
1.
Set the following parameters back to the value read before the
test.
only for Thermo-5:
Optionally, the parameter file saved before
the test can be downloaded from the USB data carrier
(
Thermo-
5 Operating instructions Chapter “Save/Load”).
Open the
Setting \ Controller
menu page.
Parameter
Measuring point internal
(Default value: “Main
line”).
Parameter
Operating mode
(Default value: “autom.”).
Open the
Settings \ Miscellaneous
menu page.
Parameter Pressure relief with unit OFF
(Default value: “ON”).
only for Series 4:
Open the
Settings \ Miscellaneous
menu page.
Parameter
Measuring point internal
(Defau
lt value: “Main
line”).
Open the
Setting \ Service \ Parameter \ Controller
menu page.
Parameter
Operating mode
(Default value: “autom.”).
Open the
Setting \ Service \ Parameter \ System
menu page.
Parameter Pressure relief with unit OFF (Default value:
“ON”).
2.
Only for units with additional equipment ZE:
Open the
Settings \ Miscellaneous
menu page.
Set the
Switch over external sensor
parameter (Default
value: “autom.”)
3.
Open the
Functions
menu page and select the
Cooling
function.
4.
Open the
Functions
menu page and select the
Mould
evacuation
function.