HB Products HBPH-MK2 Скачать руководство пользователя страница 5

 

  

 

Instructional manual- HBPH-MK2 

– UK-004 

  

  

Page 5 

 

Calibrating the Sensor   

When you receive your sensor, it is not calibrated which is needed to provide a precise result. The sensor 

calibration is made in the PLC, PC, display or were you translate your mA signal into a pH value. So, the 
calibration is in principle a formula which calculate your pH value based on the mA signal. The parameters 
in the formula are based on measurements done with two well-known solutions. Two pH buffer solutions, 
7 pH and either 4 pH or 10 pH, and a DC milliammeter are required. The calculations below are done with 
pH 7 and pH 10. The process is similar for other buffer pH values. 

 

1.

 

Disconnect the red wire at the instrument or power supply and connect it to the multimeter (-) 
black.  

2.

 

Connect the multimeter (+) red to the instrument or power supply red wire output terminal.  

3.

 

Rinse the probe and place it in 7 pH buffer solution. Allow the temperature of the buffer and probe 
to stabilize at room temperature.  

4.

 

Write down the reading on the multimeter as mA

pH7 (in the example called: ”

m

ApH7”).  The 

reading should be between 11 and 13 mA.  If not, the probe is defective, or the salt bridge needs to 
be replaced  

5.

 

Rinse the probe and place it in 10 pH buffer solution. Allow the temperature of the buffer and 
probe to stabilize at room temperature. 

6.

 

Write down the reading on the multimeter as mA

pH10 (in the example called: ”

m

ApH10”).  

 

7.

 

Reinstall  

 

The pH value is calculated using this formula: pH= K * mA  C  

 

Example: 

The two constants can be calculated like this for 7 and 10 pH solutions: 

The noted currents for the two solutions in the example are: mApH7 =11.08mA and mApH10=14.47mA 

 

K = (pH value1-pH value2) / (mApH7 - mApH10) =7 - 10 / 11.08 - 14.47 = -3/-3.39 =0.885 

C = pH value1 - K* mApH7 = 7 

 0,885 *11.08 = - 2.81 

The formula looks like this: pH = 0.885 * mA signal - 2.81 

 
Checking the Sensor   

The operation of the 2-wire, 4-20 mA sensor can be checked by a few simple measurements. Two pH buffer 

solutions, 7 pH and either 4 pH or 10 pH, and a DC milliammeter are required.   

  

Содержание HBPH-MK2

Страница 1: ...on 3 Cold Liquid applications 3 Electrical connection 4 Storage 4 Maintenance and calibration 4 Calibrating the Sensor 5 Checking the Sensor 5 Probe Cleaning 6 Replacement of Salt Bridge 7 Troubleshoo...

Страница 2: ...period thoroughly clean the sensor put a piece of cotton ball with a few drops of water into the protective cap and place it on the sensor This keeps the junction from drying out which causes slow re...

Страница 3: ...duit Do not run the power cable or control cables in the same conduit with the probe interconnect cable 3 Remove the protective plastic cap from the end of the probe before placing in service 4 The se...

Страница 4: ...measurement If you use the sensor for a precise measurement of the pH value you need frequent checks of the calibration The drift of the sensor is maximum 0 03pH per day but that is the maximum figure...

Страница 5: ...Allow the temperature of the buffer and probe to stabilize at room temperature 4 Write down the reading on the multimeter as mApH7 in the example called mApH7 The reading should be between 11 and 13 m...

Страница 6: ...ied the probe is defective or the salt bridge needs to be replaced 8 If the span of the probe drops below 2 75mA than the probe still can be used Adjustments will have to be made to the receiving devi...

Страница 7: ...th pH 7 solution or distilled water 5 Refill the chamber with a reference solution supplied with the salt bridge up to the tip of the electrode inside the chamber DO NOT OVERFILL It is important to le...

Страница 8: ...iled Sensor is non repairable The output is outside of 4 20mA The sensor electronic has failed Sensor is non repairable Parts and Accessories Description Part 3 pack salt bridge replacement kit HBPH S...

Отзывы: