Instructional manual- HBPH-MK2 

– UK-004 

  

  

Page 4 

 

 

Electrical connection                                         

HBPH probes have two wires: black and red.  The red wire is to be connected to the +24 VDC terminal and 

the black wire to the 24 VDC common terminal via the load in the loop.

     

  

  

  

  

 

 

 

Storage  

1.

 

Do not discard the protective cap that came with the sensor. If the sensor is removed from the 
process for longer periods of time, thoroughly clean the sensor, put a piece of cotton ball with a 
few drops of water into the protective cap and place it on the sensor with tape, this keeps the 
junction from drying out which causes slow response when put back into operation or causes 
permanent damage to the sensor. 

Sensors should not be left in dry lines or empty vessels for 

longer periods of time.

  

2.

 

Do not store the sensors in a dry or humid location. When storing, check the protective cap 
regularly to make sure the cotton ball remains moist. Improper storage of sensors voids the 
warranty.  

Maintenance and calibration 

The maintenance needed depends on your application and especially the accuracy required. 

Precise measurement 

If you use the sensor for a precise measurement of the pH value, you need frequent checks of the 
calibration. The drift of the sensor is maximum 0.03pH per day but that is the maximum figure for a sensor 
in high or low pH environments. Significant lower drift can be expected in neutral environments. Your 
sensor verification frequency must reflect your needed accuracy. 

Leakage detection 

When used for leakage detection the required accuracy is low and annual or 

biannual checks should be sufficient, especially when a warning is used. For 
brines with a pH value between 6 and 8 it is easy to detect ammonia. Leakage of 
1ml/l will increase the pH value from 7 to 10. Brines using NaCl or glycols will 
have a pH value like water around 7 and they can be set up with a dual level 
warning/alarm system.  

A dual level warning /alarm system contains a warning level at which you need 

to check your sensor calibration and a second level where you stop due to 
leakage. If you have a base pH value of 7 it will make sense to use a warning for 
pH above 8 and pH below 6. The second level can then be an alarm at pH=9. 
With such a system it is not necessary to make frequent checks and calibrations, 
however we recommend an annual or biannual check of the calibration.   

 

Содержание HBPH-MK2

Страница 1: ...on 3 Cold Liquid applications 3 Electrical connection 4 Storage 4 Maintenance and calibration 4 Calibrating the Sensor 5 Checking the Sensor 5 Probe Cleaning 6 Replacement of Salt Bridge 7 Troubleshoo...

Страница 2: ...period thoroughly clean the sensor put a piece of cotton ball with a few drops of water into the protective cap and place it on the sensor This keeps the junction from drying out which causes slow re...

Страница 3: ...duit Do not run the power cable or control cables in the same conduit with the probe interconnect cable 3 Remove the protective plastic cap from the end of the probe before placing in service 4 The se...

Страница 4: ...measurement If you use the sensor for a precise measurement of the pH value you need frequent checks of the calibration The drift of the sensor is maximum 0 03pH per day but that is the maximum figure...

Страница 5: ...Allow the temperature of the buffer and probe to stabilize at room temperature 4 Write down the reading on the multimeter as mApH7 in the example called mApH7 The reading should be between 11 and 13 m...

Страница 6: ...ied the probe is defective or the salt bridge needs to be replaced 8 If the span of the probe drops below 2 75mA than the probe still can be used Adjustments will have to be made to the receiving devi...

Страница 7: ...th pH 7 solution or distilled water 5 Refill the chamber with a reference solution supplied with the salt bridge up to the tip of the electrode inside the chamber DO NOT OVERFILL It is important to le...

Страница 8: ...iled Sensor is non repairable The output is outside of 4 20mA The sensor electronic has failed Sensor is non repairable Parts and Accessories Description Part 3 pack salt bridge replacement kit HBPH S...