General Information     

  

  

  

  

  

  

  

This manual covers HBPH sensor for pH measurement in brine.  

The HBPH sensor features differential design for long lifetime, user serviceability and more accurate readings. In 

typical installations these probes will last for years whereas the more common combination probe lasts only 
months.   

The sensor is a 2-wire device. It incorporates an encapsulated transmitter that outputs a 4-20 mA analog signal. It 

is designed to connect directly to a PLC.   

The output from a two-wire transmitter type is non-isolated and un-calibrated.  The system must provide 24 VDC, 

with the “low” input isolated from earth ground, and a means of calibrating for offset and span. 

 

  

NOTE

: Do not discard the protective cap that came with the sensor. If the sensor is removed from the process for 

a longer period, thoroughly clean the sensor, put a piece of cotton ball with a few drops of water into the 
protective cap and place it on the sensor.  This keeps the junction from drying out which causes slow response 
when the sensor is put back into operation or causes permanent damage to the sensor. 

Sensors should not be 

left in dry lines or empty vessels for longer periods of time.

  

Do not store the sensor in a dry or humid location. When storing, check the protective cap regularly to make sure 

the cotton ball remains moist. Improper storage of sensors voids the warranty.  

Specifications 

  

HBPH-2W-9-MKLT

 

Measurement Range  

0 to 14.00 pH 

Wetted Materials  

Ryton®(PPS), glass, ceramic, titanium, Viton 

Ambient Temperature   -15

…+50°C 

 

Liquid Temperature  

-15

…+95°C 

 

Stability  

0.03 pH/day  

Resolution  

0.01 pH  

Sensitivity  

< 0.005 pH  

Output Span  

1.14 mA/pH  

Output Offset  

12±1.0 mA @ pH 7  

Mounting  

1” NPT  

 

Max. Flow Rate  

3,0 m/sec (10 ft/sec)  

Maximum Load  

450 Ω 

 

Temperature Limits  

6,9 barg (100 psig) @ 65 °C, 2,75 barg (40 psig) @ 95°C  

Power Supply Limit  

24±4 VDC  

Cable  

4.6 m (15 ft)  

  

Содержание HBPH-MK2

Страница 1: ...on 3 Cold Liquid applications 3 Electrical connection 4 Storage 4 Maintenance and calibration 4 Calibrating the Sensor 5 Checking the Sensor 5 Probe Cleaning 6 Replacement of Salt Bridge 7 Troubleshoo...

Страница 2: ...period thoroughly clean the sensor put a piece of cotton ball with a few drops of water into the protective cap and place it on the sensor This keeps the junction from drying out which causes slow re...

Страница 3: ...duit Do not run the power cable or control cables in the same conduit with the probe interconnect cable 3 Remove the protective plastic cap from the end of the probe before placing in service 4 The se...

Страница 4: ...measurement If you use the sensor for a precise measurement of the pH value you need frequent checks of the calibration The drift of the sensor is maximum 0 03pH per day but that is the maximum figure...

Страница 5: ...Allow the temperature of the buffer and probe to stabilize at room temperature 4 Write down the reading on the multimeter as mApH7 in the example called mApH7 The reading should be between 11 and 13 m...

Страница 6: ...ied the probe is defective or the salt bridge needs to be replaced 8 If the span of the probe drops below 2 75mA than the probe still can be used Adjustments will have to be made to the receiving devi...

Страница 7: ...th pH 7 solution or distilled water 5 Refill the chamber with a reference solution supplied with the salt bridge up to the tip of the electrode inside the chamber DO NOT OVERFILL It is important to le...

Страница 8: ...iled Sensor is non repairable The output is outside of 4 20mA The sensor electronic has failed Sensor is non repairable Parts and Accessories Description Part 3 pack salt bridge replacement kit HBPH S...

Отзывы: