Table 3 Characteristic Parameter of Alarm Sound
Table 4 Occlusion response characteristics
BOON infusion set is used in this test. Test data is measured at flow rate of 1ml/h and 25ml/h
respectively, when the flow clamp is 20cm from infusion pump and the filter is 30cm from the infusion
pump,
* This infusion pump uses anti-bolus function that will be triggered after occlusion alarm, then
automatically release pressure in the tube.
Flow rate
(ml/h)
Occlusion
Alarm Level
Blockage
Pressure
(KPa)
Occlusion alarm time
Dosage
(ml)
)
1
Low
70
0h52min11sec
0.26
Medium
93
1h9min16sec
0.39
High
137
1h32min1sec
0.45
25
Low
61
0h1min45sec
0.26
Medium
97
0h2min29sec
0.42
High
147
0h3min34sec
0.50