immediate correct operation. It contains mainly photoelectric alarm (light emitting diode ) , audible
alarm (loudspeaker and buzzer) and screen display alarm;
HMI: to set parameters of infusion through keys, such as infusion volume and speed. It can display
all parameters and current working state through color LCD.
4. Technical Characteristics and Parameters
Parameters
Description
Applicable infusion set
(1) 15 drops/ml, 20 drops/ml
,
60 drops/ml infusion set
(2) Hawkmed dedicated infusion set (1ml=20 drops, optional)
Infusion accuracy
±5% (dedicated set ±3%, 10-35
℃
, test volume 2ml)
Infusion rate
0.1-1200ml/h, increment
0.01ml/h
Dripping rate
0-400 drops/min
Preset value
0-9,999.9ml, increment
0.01ml
Volume infused (∑)
36,000ml
KVO rate
(0-10)
ml/h, increment
0.01
Bolus
(0.1-1200)ml/h
Purge
1200ml/h
Occlusion pressure
10.0-160.0kpa, 8 level adjustable
Low battery alarm limit
Voltage of internal batteries is lower than 7.1V
Battery exhausted alarm limit
Voltage of internal batteries is lower than 6.65V
SPL (sound pressure level) of
alarm signal
SPL of alarm signal not lower than 45dB(A) at low grade
SPL of alarm signal not lower than 65dB (A) at high grade
Waterproof level
IPX3
AC power supply
100-240V 50/60Hz