ESC2EYEST-US-INS-LAB-RevA20 22
3.5 HEIGHT AND PATIENT SURFACE
ADJUSTMENT
3.5.11 Emergency Drop Back
This chair is equipped with a manual override
function for the back section of the chair. This
option should only be used in an emergency
situation. To activate the emergency drop
back, support the back section and pull
outward on the red activating handle located
on the patient right side under the seat to the
rear (Figure 3.5-6).
Info: When activating the emergency drop,
depending on the back angle, you may
need to push the back in order to initiate
movement.
Figure 3.5-6