Installation
CAUTION: Do not overtighten the cylinder. Overtightening the
cylinder can cause the glass cylinder to crack.
the cylinder in place.
Now that the air cylinder is in place, rotate the lead screw to position the flag pin in
between both the front and rear switches. To complete the installation, replace the top
cover of the Inspira unit. When Inspira is powered on it should display the correct
cylinder size after initializing.
Fuse Replacement
Key Information
1. Make sure the power cord is disconnected from the main supply before serv-
icing the fuse.
2. Use only Type 3AG, 1/4” x 11/4”, 1 amp, 250 volt,Type T (time delay) fuses
(Harvard Apparatus part number 5113-033 or equivalent).
Turn off power and remove power cord from power module. Use a straight blade
screwdriver to pry open the access door. Remove the fuse holder and then remove the
fuses from this holder as shown in figure 8 below. Replace fuses using only 1 amp,
250 volts,Type T fuses. Then replace the fuse holder.
Note: A fuse, which is not operator replaceable, is located on the power supply. This
fuse is rated 5 amps, 250 volts.
lead screw to the motor shaft. Turn the lead screw until the flag pin on the rear of the
plunger is vertically in-line with the rear switch as illustrated. (Note: When installing
the air cylinder, do not change the bends of the cylinder detection switches or else the
correct cylinder will not be detected.) Now insert the cylinder by placing the plunger
portion of the air cylinder on the end of Inspira’s white plunger. Carefully slide the air
cylinder back on the white plunger compressing the air cylinder just enough to allow
the front end of the cylinder to be lowered into place. Once the air cylinder is fit in
place, screw the air cylinder’s threaded nipple into the valve block and hand tighten
Figure 8. Fuse Replacement