9
H
a
rv
a
rd
A
p
p
a
ra
tu
s
I
n
s
p
ir
a
A
d
v
a
n
c
e
d
S
a
fe
ty
V
e
n
ti
la
to
r
U
s
e
r'
s
M
a
n
u
a
l
Publication 5404-001-REV-C
Installation
Electrical Connections: Front Panel
Inspira provides four BNC type connectors at the front of the machine. Their func-
tions are as follows:
Sync Out
Produces a digital pulse at the start of the respiration cycle allowing synchronization
of other electronic equipment with Inspira’s respiration cycle.
Trigger In
For use with equipment which uses a digital signal to start and stop respiration. An
example of this type of equipment would be a footswitch.
Analog Out 1
An analog signal representing the airway pressure within the ventilator. This signal
may be monitored with an oscilloscope.
Analog Out 2
Not used in this model.
Note: Electrical specifications for these signals are given in Appendix A.
Figure 4. Front Panel