7 Block OUTPUTS
Device Functions Proline Promag 55
94
Hauser
7.3
Group RELAY OUTPUT (1 to 2)
7.3.1
Function group CONFIGURATION
OUTPUTS
E
⇒
CURRENT OUTPUT 1
EAA
⇓
CURRENT OUTPUT 2
EAB
⇓
PULSE/FREQ. OUTP. 1
ECA
⇓
PULSE/FREQ. OUTP. 2
ECB
⇓
RELAY OUTPUT 1
EGA
⇒
CONFIGURATION
470
⇓
RELAY OUTPUT 2
EGB
⇒
CONFIGURATION
470
Function description
OUTPUTS
➞
RELAY OUTPUT (1 to 2)
➞
CONFIGURATION
ASSIGN RELAY
(4700)
Use this function to assign a switching function to the relay output.
Options:
OFF
ON (operation)
FAULT MESSAGE
NOTICE MESSAGE
FAULT MESSAGE or NOTICE MESSAGE
EPD (empty pipe detection, if active)
FLOW DIRECTION
LIMIT MASS FLOW
LIMIT VOLUME FLOW
LIMIT TOTALIZER (1 to 3)
LIMIT CONDUCTIVITY *
* only when conductivity is switched on
Factory setting:
FAULT MESSAGE
Advanced options with optional software package SOLID CONTENT FLOW:
LIMIT TARGET MASS FLOW
LIMIT TARGET MASS FLOW %
LIMIT TARGET VOLUME FLOW
LIMIT TARGET VOLUME FLOW %
LIMIT CARRIER MASS FLOW
LIMIT CARRIER MASS FLOW %
LIMIT CARRIER VOLUME FLOW
LIMIT CARRIER VOLUME FLOW %
Advanced options with optional software package ADVANCED DIAGNOSTICS:
LIMIT COATING E1 DEVIATION*
LIMIT COATING E2 DEVIATION*
LIMIT ELECTRODE POTENTIAL 1 DEVIATION
LIMIT ELECTRODE POTENTIAL 2 DEVIATION
LIMIT VOLUME FLOW DEVIATION
LIMIT NOISE VALUE DEVIATION
* only when coating detection is switched on
!
Note!
• It is very important to read and comply with the information on the switching
characteristics of the relay output (see Page 101).
• It is advisable to configure at least one relay output as a fault output and define the
outputs' response to error.
• The relay output is configured as a normally open (NO or make) contact by default. It
can be reconfigured as a normally closed (NC or break) contact by means of a jumper
on the relay module (see Operating Instructions Promag 55, BA119D/06).
• If you select OFF, the only function shown in the CONFIGURATION function group is
this function (4700).
Содержание Proline Promag 55
Страница 2: ...Device Functions Proline Promag 55 2 Endress Hauser ...
Страница 5: ...Device Functions Proline Promag 55 Contents Endress Hauser 5 ...
Страница 164: ...14 Index Device Functions Proline Promag 55 164 Endress Hauser ...
Страница 166: ...www endress com worldwide BA120D 06 en 10 09 71104957 FM SGML6 0 ...