7 Block OUTPUTS
Device Functions Proline Promag 55
92
Hauser
Function description
OUTPUTS
➞
PULSE/FREQUENCY OUTPUT (1 to 2)
➞
OPERATION (STATUS)
ACTUAL STATUS
(4341)
!
Note!
This function is not available unless the STATUS setting was selected in the OPERATION
MODE function (4200).
Use this function to check the actual status of the status output.
User interface:
NOT CONDUCTIVE
CONDUCTIVE
SIMULATION SWITCH
POINT
(4343)
!
Note!
This function is not available unless the STATUS setting was selected in the OPERATION
MODE function (4200).
Use this function to activate simulation of the status output.
Options:
OFF
ON
Factory setting:
OFF
!
Note!
• The "SIMULATION STATUS OUTPUT" message indicates that simulation is active.
• The measuring device continues to measure while simulation is in progress, i.e. the
actual measuring values are output correctly via the other outputs.
"
Caution!
The setting is not saved if the power supply fails.
VALUE SIMULATION
SWITCH POINT
(4343)
!
Note!
This function is not available unless STATUS was selected in the OPERATION MODE
function (4200) and the SIMULATION SWITCH POINT function (4343) is active
(= ON).
Use this function to define the switching response of the status output during the
simulation. This value is used to test downstream devices and the flowmeter itself.
Options:
NOT CONDUCTIVE
CONDUCTIVE
Factory setting:
NOT CONDUCTIVE
"
Caution!
The setting is not saved if the power supply fails.
Содержание Proline Promag 55
Страница 2: ...Device Functions Proline Promag 55 2 Endress Hauser ...
Страница 5: ...Device Functions Proline Promag 55 Contents Endress Hauser 5 ...
Страница 164: ...14 Index Device Functions Proline Promag 55 164 Endress Hauser ...
Страница 166: ...www endress com worldwide BA120D 06 en 10 09 71104957 FM SGML6 0 ...