Device Functions Proline Promag 55
8 Block INPUTS
Hauser
109
VALUE 20 mA
(5203)
Use this function to assign a value to the 20 mA current.
User input:
5-digit floating-point number
Factory setting:
Dependent on the process variable assigned to the current input (see function
ASSIGN CURRENT, 5200).
– Density: 2.0 kg/l
– Temperature: 200 °C
!
Note!
The appropriate unit is taken from the functions UNIT DENSITY (0420) or
UNIT TEMPERATURE (0422).
ERROR-VALUE
(5204)
Use this function to enter a defined error value for the process variable in question.
If the current value lies outside of the selected range (see function CURRENT SPAN,
5201), then the process variable is set to the "error value" defined here and a
corresponding notice message CURRENT INPUT RANGE (# 363) is generated.
User input:
5-digit floating-point number
Factory setting:
Dependent on the process variable assigned to the current input (see function
ASSIGN CURRENT, 5200).
– Density: 1.25 kg/l
– Temperature: 75 °C
!
Note!
• Triggered amplifier faults or the error behavior of the outputs do not have any
influence on the current input.
• The appropriate unit is taken from the functions UNIT DENSITY (0420) or
UNIT TEMPERATURE (0422).
Function description
INPUTS
➞
CURRENT INPUT
➞
CONFIGURATION
Содержание Proline Promag 55
Страница 2: ...Device Functions Proline Promag 55 2 Endress Hauser ...
Страница 5: ...Device Functions Proline Promag 55 Contents Endress Hauser 5 ...
Страница 164: ...14 Index Device Functions Proline Promag 55 164 Endress Hauser ...
Страница 166: ...www endress com worldwide BA120D 06 en 10 09 71104957 FM SGML6 0 ...