Bit 0
Not used.
Bit 1
Not used.
Bit 2
Not used.
Bit 3
Not used.
Bit 4
Measurement invalid bit.
Bit 5
Not used.
Bit 6
Not used.
Bit 7
Not used.
Example: STAT:QUES:COND?
Response: 16
8.4.11.23
STATus:QUEStionable:ENABle?
This command returns the Questionable Status Enable Register (see STA-
Tus:QUEStionable:ENABle <num> command below).
Example: STAT:QUES:ENAB?
Response: 16
8.4.11.24
STATus:QUEStionable:ENABle <num>|MIN|MAX|DEF
This command sets the Questionable Status Enable Register. This register de-
termines which event bits of the Questionable Status Event Register affect the
Questionable Status Bit of the Status Byte Register. If any bit in the Question-
able Status Event Register is set (1) while its corresponding mask bit in the
Questionable Status Enable Register is set, the Questionable Status Bit in the
Status Byte Register will be set. The value of the <num> parameter is a number
from 0 to 15 that is the sum of the binary-weighted values of each mask bit.
The Questionable Status Event Register is described in Section 8.4.11.21, STA-
Tus:QUEStionable?. The following example causes the Questionable Status Bit
in the Status Byte Register to be set when a measurement is invalid.
Example: STAT:QUES:ENAB 16
Response: 16
8.4.12
Port Numbers
The following table lists the communication ports and their number setting.
Number
Port
0
Serial (default)
136
Hart Scientific
8 Digital Communications Interface