8.4.8.12
SYSTem:COMMunicate:SERial:TIMe <num>|MIN|MAX|DEF
Sets to serial port sample period setting in seconds.
Example: SYST:COMM:SER:TIM 120
The <num> parameter can be one of the following values: 1, 2, 5, 10, 15, 30,
60, 120, 300, 600, 900, 1200, 1800, or 3600. If a value other than one listed
here is used, the instrument automatically selects the nearest acceptable value
that is less than the value entered. Values greater than 3600 or less than 1 are
ignored. Specifying the MIN, MAX or DEF parameter sets the sample period
setting to the minimum, maximum, or default value respectively.
8.4.8.13
SYSTem:KLOCkout?
Returns the state of the button panel lockout.
Example: SYST:KLOC?
Response: 0
A value of 1 is returned if the button panel is locked out. A value of 0 is re-
turned if the button panel is enabled (unlocked).
8.4.8.14
SYSTem:KLOCkout <bool>
Sets the state of the button panel lockout.
Example: SYST:KLOC 1
The <bool> parameter turns the button panel lockout on (1 or ON) or off (0 or
OFF)..
8.4.8.15
SYSTem:POWer:BATTery?
Returns the current battery charge level.
Example: SYST:POW:BATT?
Response: 90
The returned value indicates the current percentage of charge remaining in the
battery. The battery charge level is approximate and should be used only as a
guideline.
8.4.9
Date and Time Commands
The date and time commands are used for accessing the date and time related
settings.
1620 the “DewK”
119
8 Digital Communications Interface