Instrument Configuration
73
•
•
•
•
•
•
On the
HI 6500 series instrument display
, select the calibration menu and use the UP or
DOWN button to select the C2 eCAL submenu, then select the Ref Weight menu item. The
current value will be displayed, press the ENTER or the RIGHT button if this value needs
to be modified.
Sensitivity Parameter
The load cell sensitivity, defined in mV/V, sets the expected change in analog signal over
the full scale range of an analog strain gage load cell. For example if a 5 Volt excitation is
applied to a load cell with a sensitivity of 2 mV/V then the full scale signal will be 10mV.
To ensure that the HI 6500 optimizes the processing of the analog signal from the load cell,
the load sensitivity parameter should be set to match that of the load cell.
On the
Web page
, select the calibration menu and using the Sensitivity drop down list
select the required sensitivity value.
On the
HI 6500 series instrument display
, select the calibration menu and use the UP or
DOWN button to select the C2 eCAL submenu, then select the Sensitivity menu item. The
list of sensitivity values supported will be displayed and the current value will be
highlighted, if the sensitivity needs to be modified use the UP and DOWN buttons to select
a new value and press the ENTER button.
Hard Cal Commands and Parameters
These commands and parameters are used when doing a Hard Cal or traditional calibration
of the instrument.
Cal Lo Weight Parameter
The value you enter for Cal Lo Weight sets the lower weight value to be used when
calibrating the system. The Cal Lo weight value, is normally zero (no weight on the scale),
but can be set to any known weight on the scale. With nothing (zero) on the scale, the Cal
Lo Weight is 0.00. With 5 lbs on the scale, the Cal Lo Weight is 5.00 lbs.
Содержание HI 6500 Series
Страница 10: ...viii Contents ...
Страница 20: ...10 Chapter 2 ...
Страница 72: ...62 Chapter 4 ...
Страница 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Страница 97: ...Instrument Configuration 87 Serial Option Termination ...
Страница 126: ...116 Chapter 5 ...
Страница 156: ...146 Chapter 8 ...
Страница 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Страница 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Страница 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Страница 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Страница 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Страница 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Страница 169: ...Troubleshooting 159 B3 Guidelines to Verify Configuration Filter Settings to Improve Stability ...
Страница 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Страница 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Страница 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Страница 174: ...164 Chapter 9 H Mechanical Installation ...
Страница 175: ...Troubleshooting 165 J Electrical Inspection ...
Страница 176: ...166 Chapter 9 K Installation Check Points ...
Страница 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Страница 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Страница 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Страница 195: ...Installation of Swivel Mount Enclosures 185 Figure D2 Assembling the Swivel Mount Enclosure on a Table ...
Страница 196: ...186 Appendix A Dimensions ...
Страница 200: ...4 HI 6500 Series USER GUIDE Index ...
Страница 201: ......