124
Chapter
6
•
•
•
•
•
•
Step 4. Click the Do Cal Lo button to do the hard Cal Zero operation. A “Cal Completed
OK” message appears briefly if the calibration was successful. An Error number
appears if the calibration was not successful. See the Error list in the
troubleshooting chapter for help in correcting the error.
Step 5. To set the Span Weight, place certified test weights on the scale.
Step 6. In the Span Weight text field, enter the amount of test weight that was placed on
the scale.
Step 7. Click the Do Cal Hi button. A “Cal Completed OK” message appears briefly if
the calibration was successful.
Step 8. An Error message appears if the calibration was not successful. See the error list
in the Troubleshooting chapter, for help in correcting errors.
Step 9. The scale is now calibrated.
Hard Calibration from the Front Panel
Step 1. From the Configuration menu, use the down arrow to select Calibration, and then
press Enter to display the Calibration menu. Select the Hard Cal menu item and
press the ENTER button to open the menu.
Step 2. Use the UP or DOWN button to select the Ref Weight value then press the
ENTER to set the reference weight equal to the weight on the scale. Normally,
you would remove all weight "live load" from the scale to obtain a Ref Weight of
0.0. Wait 12 seconds or more for the reading to stabilize. Press the ENTER button
to save the value and return to calibration menu.
Step 3. Use the DOWN button to select " Do Cal Lo, and press the ENTER button to set
the low calibration point. A "Cal Completed OK" message appears briefly if the
calibration was successful.
Step 4. An Error appears if the calibration was not successful. See the Troubleshooting
chapter, for help with errors.
Step 5. Press the DOWN button to select Span Weight, the current span weight will be
displayed. Press the ENTER button to modify the Span Weight value.
Содержание HI 6500 Series
Страница 10: ...viii Contents ...
Страница 20: ...10 Chapter 2 ...
Страница 72: ...62 Chapter 4 ...
Страница 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Страница 97: ...Instrument Configuration 87 Serial Option Termination ...
Страница 126: ...116 Chapter 5 ...
Страница 156: ...146 Chapter 8 ...
Страница 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Страница 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Страница 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Страница 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Страница 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Страница 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Страница 169: ...Troubleshooting 159 B3 Guidelines to Verify Configuration Filter Settings to Improve Stability ...
Страница 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Страница 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Страница 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Страница 174: ...164 Chapter 9 H Mechanical Installation ...
Страница 175: ...Troubleshooting 165 J Electrical Inspection ...
Страница 176: ...166 Chapter 9 K Installation Check Points ...
Страница 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Страница 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Страница 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Страница 195: ...Installation of Swivel Mount Enclosures 185 Figure D2 Assembling the Swivel Mount Enclosure on a Table ...
Страница 196: ...186 Appendix A Dimensions ...
Страница 200: ...4 HI 6500 Series USER GUIDE Index ...
Страница 201: ......