background image

INSTRUCTION MANUAL

HI98107

Pocket-sized pH Meter

Warranty

This meter is warranted for a period of one year against defects in workmanship 

and materials when used for their intended purpose and maintained according to 

instructions. The electrode is warranted for a period of six months. This warranty is 

limited to repair or replacement free of charge. Damage due to accidents, misuse, 

tampering or lack of prescribed maintenance is not covered. If service is required, 

contact your local Hanna Instruments Office. If under warranty, report the model 

number, date of purchase, serial number and the nature of the problem.

If the repair is not covered by the warranty, you will be notified of the charges incurred. 

If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods 

Authorization (RGA) number from the Technical Service department and then send it 

with shipping costs prepaid. When shipping any instrument, make sure it is properly 

packaged for complete protection.

IST98107N   06/18

Accessories

Electrode Storage Solution

Code

Description

HI70300M

Electrode storage solution, 230 mL bottle

Electrode Cleaning Solution

Code

Description

HI7061M

General purpose cleaning solution, 230 mL bottle

pH Buffer Solution

Code

Description

HI70004P

pH 4.01 buffer solution, 20 mL sachets (25 pcs.)

HI70007P

pH 7.01 buffer solution, 20 mL sachets (25 pcs.)

HI70010P

pH 10.01 buffer solution, 20 mL sachets (25 pcs.)

HI77400P

pH 4.01 & 7.01 buffer solution, 20 mL sachets  
(10 pcs., 5 ea.)

HI770710P

pH 10.01 & 7.01 buffer solution, 20 mL sachets  
(10 pcs., 5 ea.)

Care and Maintenance

To obtain the highest accuracy for measurements it is important to follow these tips:

• Calibration is only as good as the buffers being used. The pH buffers value change 

over time once the sachets are opened. Fresh buffer should be used for each 

calibration.

• The electrode should be rinsed with purified water each time before placing in 

buffer or sample to be tested.

• When the meter is not in use it is important to add several drops of storage solution 

to a sponge in the protective cap to keep the electrode hydrated. If storage 

solution is not available, then pH 4.01 or pH 7.01 buffer can be used.

• For improved accuracy it is recommended to calibrate in two buffers. It is important 

to use buffers that bracket the expected value of the sample to be tested. For 

example, if the expected value is pH 8, the meter should be calibrated using  

pH 7.01 and pH 10.01 buffers.

Battery Replacement

The meter features a low battery indicator. When the battery is running low (under 

10%), the battery indicator will blink on the LCD. When the battery is discharged  

dEAd bAtt

” will be displayed on the LCD for 2 seconds and the meter will turn off.

To change the CR2032 Li-ion battery, turn the battery cover located on the back of 

the meter counterclockwise to unlock. Remove cover and replace with new battery 

+ side facing up.

Note: Batteries should only be replaced in a safe area using the battery 

type specified in this instruction manual. Old batteries should be disposed in 

accordance with local regulations.

07

07

• Inorganic 

Soak in Hanna Instruments HI7074 

 

Inorganic Cleaning Solution for 15 minutes.

• Oil and grease 

Rinse with Hanna Instruments HI7077

 

 

Oil and grease Cleaning Solution

IMPORTANT: After performing any of the cleaning procedures rinse the electrode 

thoroughly with distilled water, and soak the electrode in HI70300 Storage Solution 

for at least 1 hour before taking measurements.

• Removal of films, dirt or deposits on the membrane/junction:
• General 

Soak in Hanna Instruments HI7061 

 

General Cleaning Solution for approximately 1 hour.

• Protein 

Soak in Hanna Instruments HI7073 

 

Protein Cleaning Solution for 15 minutes.

• If the electrode or junction is dirty soak the tip in HI7061 cleaning solution for 

30 minutes, rinse thoroughly in distilled water and then follow cleaning procedure.

Two-point calibration

Proceed with steps 1 through 3 under single-point calibration using pH 7.01 buffer 

first. Then follow steps below:

• The “pH 4.01 USE“ message is then displayed.
• Place the electrode in the second calibration buffer (pH 4.01 or 10.01). When the 

second buffer is accepted, the LCD will display “Stor” for 1 second and the meter 

will return to the normal measurement mode.

• The “CAL“ tag will be displayed in measurement mode with the calibrated buffers.
• If the buffer is not recognized or the slope is out of accepted range “---- Err“ 

is displayed. Change the buffer, clean the electrode or press any key to exit 

calibration.

It is always recommended to carry out a two-point calibration for better accuracy.

Clear calibration

Press CAL button. Meter enters in calibration mode.
Press ON/OFF button and CLR is displayed. The meter will now be at default calibration.
The “CAL“ tag will be cleared in measurement mode, indicating the missing calibration, 

until a new calibration will be performed.

Error messages

• During user calibration, if the reading is out of the accepted range, the meter will 

displayed “---- Err“.

• pH 14.0 or 0.0 blinking means the limit of the measured value is out of range. 

Verify that the electrode is in solution.

• If the measured temperature is higher than 50.0 ºC or lower than 0.0 ºC, the 

50.0 ºC or 0.0 

°

C message is displayed blinking.

• If the buffer is not recognised “----Err“ is displayed. Check the buffer, clean the 

electrode or press CAL button to exit calibration.

• In case of erroneous readings even after an accurate conditioning and calibration, 

the reference junction might be contaminated or clogged. Pull out 2 mm (1/8’’) of 

the cloth junction to renew the electrode reference (it is recommended to cut the 

cloth leaving always at least 2 mm - (1/8’’) over the reference compartment) and 

recalibrate the meter. The cloth junction can be pulled out approximately 20 times. 

After that, the electrode will have to be replaced.

pHep

Содержание pHep

Страница 1: ...INSTRUCTION MANUAL HI98107 Pocket sized pH Meter pHep ...

Страница 2: ...e remove the protective cap and condition the electrode by soaking the tip bottom 3 cm 1 18 in HI70300 storage solution or in pH 7 01 buffer solution for several hours Then follow the calibration procedure Do not be alarmed if white crystals appear around the cap This is normal with pH electrodes and they dissolve when rinsed with water Turn the pH electrode on by pressing ON OFF button Remove the...

Страница 3: ... 1 pH 0 5 C 1 0 F Calibration automatic one or two points pH 4 01 7 01 10 01 Temperature Compensation automatic 0 to 50 C Battery Type CR2032 3V Li ion 1 pc Battery Life approximately 800 hours of continuous use Environment 0 to 50 C 32 to 122 F RH 100 max Dimensions 160 x 40 x 17 mm 6 3 x 1 6 x 0 7 Weight without battery 65 g 2 3 oz Specifications Recommended Immersion Level ...

Страница 4: ...compartment The meter will enter in setup mode Press the ON OFF button to move through setup parameters Press the CAL button to change the settings All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner for another second The meter then enters the normal measuring mode Note Keeping the ON OFF button pressed while turning the meter ...

Страница 5: ...ode tip thoroughly to eliminate cross contamination After cleaning rinse the electrode tip with some deionized water and some of the sample to be measured pH calibration Enter calibration mode while in pH measurement mode Place the electrode into the first calibration buffer If performing a two point calibration use pH 7 01 buffer first The meter will enter the calibration mode displaying pH 7 01 ...

Страница 6: ...lectrode in the second calibration buffer pH 4 01 or 10 01 When the second buffer is accepted the LCD will display Stor for 1 second and the meter will return to the normal measurement mode The CAL tag will be displayed in measurement mode with the calibrated buffers If the buffer is not recognized or the slope is out of accepted range Err is displayed Change the buffer clean the electrode or pres...

Страница 7: ...any of the cleaning procedures rinse the electrode thoroughly with distilled water and soak the electrode in HI70300 Storage Solution for at least 1 hour before taking measurements Removal of films dirt or deposits on the membrane junction General Soak in Hanna Instruments HI7061 General Cleaning Solution for approximately 1 hour Protein Soak in Hanna Instruments HI7073 Protein Cleaning Solution f...

Страница 8: ...charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization RGA number from the Technical Service department and then send it with shipping costs prepaid When shipping any instrument make sure it is properly packaged for complete protection Accessories Electrode Storage Solution Code Description HI70300M Electrode storage solution 230 mL bo...

Отзывы: