Troubleshooting
• In calibration mode, the “--- Err” message is displayed if the reading is
outside accepted range or buffer solution is not recognized.
• If the probe is not in buffer solution, “pH 14.00” or “pH 0.00” is displayed
blinking.
• If measured temperature is outside accepted range, “0
°
C” or “50
°
C” is
displayed blinking.
Care & Maintenance
Please read the information below to ensure highest possible accuracy.
• Fresh buffers should be used for each calibration, once the sachets are open
the buffer value can change over time.
• If the measurements are taken sucessively, rinse the probe thoroughly in
purified water to eliminate cross-contamination.
• When not in use, add a few drops of storage solution to the protective cap to
keep the junction hydrated. If storage solution is not available, pH 4.01 or pH
7.01 buffer can be used.
• For improved accuracy, a two-point calibration is recommended.
• When dirty, the junction can be pulled out to expose a fresh section. Pull
out 3 mm (1/8”) of the cloth, making sure that the reference compartment
remains covered. This procedure can be repeated approximately 9 times,
renewing the tester’s life.
• If the electrode or junction is dirty, soak the tip in
HI7061
General purpose
cleaning solution for 30 minutes. Rinse with distilled water and follow
cleaning procedure.
Film deposits and other impurities that accumulate with use must be removed
periodically off the membrane or junction.
•
General:
Soak in
HI7061
General cleaning solution for approximately 1 hour.
•
Protein:
Soak in
HI7073
Protein cleaning solution for 15 minutes.
•
Inorganic:
Soak in
HI7074
Inorganic cleaning solution for 15 minutes.
•
Oil & grease:
Rinse with
HI7077
Oil and grease cleaning solution.
Note:
After performing any of the cleaning procedures rinse the electrode
thoroughly with distilled water, and soak in
HI70300
Storage solution for
at least 1 hour before taking measurements.
Battery Replacement
Battery life percentage indicator is displayed at power on. If the battery
level drops below 10%, the battery symbol starts blinking. If the battery is
discharged, “dEAd bAtt” is displayed and then the tester turns off.
08
08
To change the CR2032 Lithium-ion battery, turn the battery cover located on the
back of the tester counterclockwise to unlock. Remove the cover and replace the
battery with positive (+) side facing out.
Note:
Only use the battery type specified in the manual. Old batteries
should be disposed in accordance with local regulations.
Accessories
pH Buffer Solution
HI70004P
pH 4.01 buffer solution, 20 mL sachet (25 pcs.)
HI70007P
pH 7.01 buffer solution, 20 mL sachet (25 pcs.)
HI70010P
pH 10.01 buffer solution, 20 mL sachet (25 pcs.)
HI77400P
pH 4.01 & 7.01 buffer solution, 20 mL sachet (10 pcs., 5 each)
HI770710P
pH 10.01 & 7.01 buffer solution, 20 mL sachet (10 pcs., 5 each)
Electrode Cleaning Solution
HI7061M
General purpose cleaning solution, 230 mL
HI7073
Protein cleaning solution
HI7074
Inorganic cleaning solution
HI7077
Oil and Grease cleaning solution
Electrode Storage Solution
HI70300M
Electrode storage solution, 230 mL
Certification
All Hanna Instruments conform to the
CE European Directives.
Disposal of Electrical & Electronic Equipment.
The product
should not be treated as household waste. Instead hand it over
to the appropriate collection point for the recycling of electrical
and electronic equipment which will conserve natural resources.
Disposal of waste batteries.
This product contains batteries, do
not dispose of them with other household waste. Hand them over
to the appropriate collection point for recycling.
Ensuring proper product and battery disposal prevents potential negative
consequences for the environment and human health. For more information,
contact your city, your local household waste disposal service, the place of
purchase or go to www.hannainst.com.
Recommendations for Users
Before using this product, make sure it is entirely suitable for your specific
application and for the environment in which it is used. Avoid touching the
electrode area. Any variation introduced by the user to the supplied equipment
may degrade the tester’s performance. For yours and the tester’s safety do not
use or store the tester in hazardous environments.
Warranty
This tester is warranted for a period of one year against defects in workmanship
and materials when used for their intended purpose and maintained according
to instructions. This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed maintenance
is not covered. If service is required, contact your local Hanna Instruments
Office. If under warranty, report the model number, date of purchase, serial
number and the nature of the problem.
If the repair is not covered by the warranty, you will be notified of the charges
incurred. If the instrument is to be returned to Hanna Instruments, first obtain
a Returned Goods Authorization (RGA) number from the Technical Service
department and then send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
Hanna Instruments reserves the right to modify the design, construction, or
appearance of its products without advance notice.
All rights are reserved. Reproduction in whole or in part is prohibited without the written
consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island,
02895, USA.
IST98108N 12/19
RoHS
compliant
HI98108
Pocket-sized pH Tester
INSTRUCTION MANUAL