Hanna Instruments HI99151 Скачать руководство пользователя страница 2

pH CALIBRATION PROCEDURE

Calibration messages:

USE : 

default suggested buffer.

REC :

 a valid buffer is recognized.

WRNG :

 invalid buffer detected.

Two-point procedure

:

• Place the probe in pH 7.01 (or 6.86) buffer.
• After the first point is accepted, the “

pH 4.01 USE

” message appears.

• Place the probe in pH 4.01, 10.01 (or 9.18) buffer.
• When the second point is accepted, the “

OK 2

” message appears and then

the meter returns to normal mode.

Single-point procedure

:

With pH 4.01, 9.18 or 10.01 buffer:

 when the calibration point is

accepted, the buffer value and “

OK 1

” message appear and then the meter

returns to normal mode.

With pH 7.01 or 6.86 buffer:

 after the calibration point has been

accepted, the 

ON/OFF/MODE

 button must be pressed to return to normal

mode.

NOTES

Probe

FC 214D

 pH electrode with internal temperature sensor, titanium

body, for Beer Industry, with DIN connector.

• If the 

probe is dry

, soak it in 

HI 70300

 Storage Solution for half an hour.

To clean the meter:

 use water.

To quit calibration (ESC):

 press 

ON/OFF/MODE

 after entering the

calibration mode and before the first point is accepted.

To clear calibration (CLR):

 press 

SET/HOLD

 after entering the calibration

mode and before the first point is accepted.

• When the 

battery

 symbol 

-    +

 appears, the batteries should be replaced

soon.

• Open the battery compartment cap on the bottom of the instrument and

replace old batteries with new ones, following the instructions on the rear of
the instrument.