23
METHOD PROCEDURE
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
• Insert the reacted sample (#2) into the holder
and ensure that the notch on the cap is positioned
securely in the groove.
• Press
Read
to start the reading. The instrument displays the result in
mg/L
as
SDBS.
INTERFERENCES
Interference may be caused by:
• Absorption particulate matter, Cationic surfactants, Strong oxidants (Cl
2
, H
2
O
2
,
S
2
O
8
2
- etc.), Sulfide cause negative interference
• Organic sulfates, Sulfonates cause positive interference
• Highly buffered samples or with extreme pH may exceed the buffering capacity of
the reagent, pH should be adjusted between 4 and 9 with diluted NaOH or HCl prior
to addition of the reagent
#2