21
METHOD PROCEDURE
• Invert the vial twice and remove the cap to release any
pressure that has built up.
• Replace the cap and shake it vigorously for 30 seconds.
Note:
Ensure the cap is secure when shaking.
• Press
Zero
and the display will show a 2 minutes countdown. To skip the timer, press
Zero
twice. To stop the timer press
Stop
or to start the 2 minute timer press
2:00
min
. During this period the chloroform layer separates from the aqueous layer, the
color of the aqueous layer will fade slightly, while the chloroform layer will turn blue.
• Remove the cap.
• Remove the upper aqueous layer using the long plastic
pipette, do not remove the lower chloroform layer.
• Add 15 mL of deionized water to the vial (up to the
25 mL mark).
• Add 2 drops of
HI95769A-0
Anionic Surfactants
Reagent A.
• Invert the vial twice and remove the cap to release any
pressure that has built up.
• Replace the cap and shake it vigorously for 30 seconds.
Note:
Ensure the cap is secure when shaking.
×2