21
METHOD PROCEDURE
• Insert the cuvette into the holder and ensure
that the notch on the cap is positioned securely
in the groove.
• Press
Read
. The display will show a 3 minute countdown prior to the measurement.
To skip the timer, press
Read
twice
.
When the timer ends, the meter will perform the
reading. The instrument displays concentration in
mg/L
of
iron (Fe).
INTERFERENCES
Interference may be caused by:
• Chloride above 185000 mg/L
• Magnesium above 100000 mg/L CaCO
3
• Calcium above 10000 mg/L CaCO
3
• Molybdate Molybdenum above 50 mg/L