20
METHOD PROCEDURE
8. METHOD PROCEDURE
REQUIRED REAGENTS
Code Description Quantity
HI93721-0
Iron HR Reagent
1 packet
REAGENT SETS
HI93721-01
Iron HR Reagent - 100 tests
HI93721-03
Iron HR Reagent - 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
Note:
If tutorial mode is disabled, follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill the cuvette with 10 mL of unreacted sample (up
to the mark). Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette and add the content of
one packet of
HI93721-0
Iron HR Reagent.
Replace the plastic stopper and the cap. Shake
gently until completely dissolved.
10 mL