Specifications
4
3. SPECIFICATIONS
Chemical Oxygen Demand LR
Range
0 to 150 mg/L (as O
2
)
Resolution
1 mg/L
Accuracy
±5 mg/L or ±4 % of reading at 25 °C, whichever is greater
Method
Adaptation of the US EPA 410.4 Approved Method for the COD Determination on
Surface Waters and Wastewaters
LED
420 nm
Chemical Oxygen Demand MR
Range
0 to 1500 mg/L (as O
2
)
Resolution
1 mg/L
Accuracy
±15 mg/L or ±4 % of reading at 25 °C, whichever is greater
Method
Adaptation of the US EPA 410.4 Approved Method for the COD Determination on
Surface Waters and Wastewaters
LED
610 nm
Note
: Range is reduced to 1000 mg/L (as O
2
) when HI93754G-25 reagents are used.
Chemical Oxygen Demand HR
Range
0 to 15000 mg/L (as O
2
)
Resolution
1 mg/L
Accuracy
±150 mg/L or ±2 % of reading at 25 °C, whichever is greater
Method
Adaptation of the US EPA 410.4 Approved Method for the COD Determination on
Surface Waters and Wastewaters
LED
610 nm
Chemical Oxygen Demand UHR
Range
0 to 60.0 g/L (as O
2
)
Resolution
0.1 g/L
Accuracy
±0.5 g/L ±3 % of reading at 25 °C
Method
Adaptation of the US EPA 410.4 Approved Method for the COD Determination on
Surface Waters and Wastewaters
LED
610 nm