20
The reading should be within specifications as reported in the
CAL Check
TM
Standard Certificate. If
the value is found out of the specifications, please check that the cuvettes are free of fingerprints,
oil or dirt and repeat validation. If results are still found out of specifications, then recalibrate the
instrument.
• Remove the cuvette.
• Place the specific
CAL Check
TM
Standard Cuvette B into the cuvette holder, for:
Free Chlorine:
B
,
HI96701‑11
Total Chlorine:
B
,
HI96711‑11
Cyanuric Acid: B,
HI96722‑11
pH: B,
HI96710‑11
Ensure that the notch on the cap is positioned securely into the groove.
• Press
CAL Check
TM
and the lamp, cuvette and detector icons together with
“CAL CHECK”
will
appear on the display, depending on the measurement phase.
• At the end of the measurement the display will show the validation standard value.
VALID
ATION PROCEDURE