18
Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:
Do not validate the instrument with any standard solutions other than the Hanna
Intruments
CAL Check
TM
Standards, otherwise erroneous results will be obtained. For accurate
validation, please perform test at room temperature: 18 to 25 °C (64.5 to 77.0 °F)
• Turn the meter on by pressing
ON/OFF
.
• When the beeper sounds briefly and the LCD displays dashes, the meter is ready.
• Place the
CAL Check
TM
Standard
HI96707‑11
Cuvette A into the cuvette holder and ensure
that the notch on the cap is positioned securely into the groove.
• Press
ZERO/CFM
and the lamp, cuvette and detector icons will appear on the display, depending
on the measurement phase.
VALID
ATION PROCEDURE