background image

HI83141-1

Portable 

pH / mV / Temperature 

Meter

INSTRUCTION MANUAL

7. Wait a few minutes and use the calibration screwdriver to adjust the 

OFFSET trimmer until it displays the 

pH value at the previously noted 

temperature. See pH Buffer Values 

at Various Temperatures table.

8.  Press the 

pH

 key.

9.  Rinse with second buffer being used for calibration (e.g. pH 4.01 

or pH 10.01). Place the tip of the electrode 4 cm (1 

1/2

”) into the 

second buffer. Stir gently.

10. Press 

TEMP

 key to read buffer temperature. Take a note of the 

displayed value.

11. Press 

pH

 key to take a pH reading.

12. Wait a couple of minutes and 

adjust the SLOPE trimmer until 
it displays the pH value at the 
previously noted temperature. 
See pH Buffer Values at Various 
Temperatures table.

13. Press 

CAL

 key. The pH calibration is now complete.

Measurement

1.  Turn the meter on.
2. After initialization, battery status is displayed. 

Replace the battery if BATT tag is displayed 
blinking.

3.  Connect the probe (pH or ORP) to the meter. 
4.  For accurate pH measurements, calibrate the 

meter for pH before use.

5.  Place the tip of the electrodes 4 cm (1 

1/2

”) into the sample.

6.  Press the designated key to select corresponding measurement 

mode (i.e. 

pH

 for pH measurements, 

mV

 for mV or ORP 

measurements, and 

TEMP

 for temperature measurements).

pH

 

mV

 

 

TEMP

7.  Stir briefly and wait a few minutes for the measurement to 

stabilize. Measured value and unit tag is displayed for the 
selected parameter.

Note: 

If measurements are taken in different samples successively, 

rinse the probes between measurements to avoid cross 

contamination. After cleaning, rinse the probes with some of the 

samples to be measured next.

Warnings

If measured value is outside the parameter limit of the probe, the 
maximum or minimum value is displayed blinking.

OFFSET

SLOPE

4 cm (1

½

”)

pH Buffer Values at Various Temperatures

Temp

pH Values

°C

°F

4.01

7.01

10.01

0

32

4.01

7.13

10.32

5

41

4.00

7.10

10.24

10

50

4.00

7.07

10.18

15

59

4.00

7.04

10.12

20

68

4.00

7.03

10.06

25

77

4.01

7.01

10.01

30

86

4.02

7.00

9.96

35

95

4.03

6.99

9.92

40

104

4.04

6.98

9.88

45

113

4.05

6.98

9.85

50

122

4.06

6.98

9.82

55

131

4.07

6.98

9.79

60

140

4.09

6.98

9.77

65

149

4.11

6.99

9.76

70

158

4.12

6.99

9.75

75

167

4.14

7.00

9.74

80

176

4.16

7.01

9.73

85

185

4.17

7.02

9.74

90

194

4.19

7.03

9.75

95

203

4.20

7.04

9.76

Accessories

Ordering info.Product description

HI1230B

Plastic body, double-junction, gel, combination pH 

electrode with BNC connector and 1 m (3.28 ft) cable

HI3131B

Glass-body, refillable, combination platinum ORP 

electrode with BNC connector and 1 m (3.28 ft) cable

HI7662

Temperature probe with 1 m (3.28 ft) cable

HI7004M

pH 4.01 buffer solution, 230 mL

HI7006M

pH 6.86 buffer solution, 250 ml

HI7007M

pH 7.01 buffer solution, 230 mL

HI7009M

pH 9.18 buffer solution, 250 ml

HI7010M

pH 10.01 buffer solution, 230 mL

HI70300M

Storage solution, 230 mL

HI7061M

General cleaning solution, 230 mL

HI7091L

Reducing pretreatment solution, 
500 mL + 14 g (set)

HI7092M

Oxidizing pretreatment solution, 250 mL

HI731326

Calibration screwdriver (20 pcs.)

HI76405

Electrode holder

Certification

All Hanna

®

 instruments conform to the 

CE European Directives.

  

  

Disposal of Electrical & Electronic Equipment.

 The product should 

not be treated as household waste. Instead, hand it over to the 
appropriate collection point for the recycling of electrical and electronic 
equipment, which will conserve natural resources.

Disposal of waste batteries.

 This product contains batteries, do not 

dispose of them with other household waste. Hand them over to the 
appropriate collection point for recycling.
Ensuring proper product and battery disposal prevents potential 
negative consequences for the environment and human health. For 
more information, contact your city, your local household waste 
disposal service, or the place of purchase.

Recommendations for Users

Before using this meter, make sure that it is entirely suitable for 
your specific application and for the environment in which it is used. 
Any variation introduced by the user to the supplied equipment may 
degrade the meter’s performance. For your and the meter’s safety do 
not use or store the meter in hazardous environments.

Warranty

HI83141-1

 is warranted for a period of two years against defects 

in workmanship and materials when used for its intended purpose 
and maintained according to instructions. Electrodes and probes 
are warranted for a period of six months. This warranty is limited 
to repair or replacement free of charge. Damage due to accidents, 
misuse, tampering, or lack of prescribed maintenance is not covered. 
If service is required, contact your local Hanna Instruments

®

 office. 

If under warranty, report the model number, date of purchase, serial 
number and the nature of the problem. If the repair is not covered 
by the warranty, you will be notified of the charges incurred. If the 
instrument is to be returned to Hanna Instruments office, first obtain 
a Returned Goods Authorization (RGA) number from the Technical 
Service department and then send it with shipping costs prepaid. 
When shipping any instrument, make sure it is properly packaged for 
complete protection.

Hanna Instruments reserves the right to modify the design, 

construction, or appearance of its products without advance notice.

All rights are reserved. Reproduction in whole or in part is prohibited without 

the written consent of the copyright owner, Hanna Instruments Inc., 

Woonsocket, Rhode Island, 02895, USA.

IST83141-1 01/22

Содержание HI83141-1

Страница 1: ...HI83141 1 Portable pH mV Temperature Meter INSTRUCTION MANUAL ...

Страница 2: ...I3131B refillable ORP electrode not supplied with the product for accurate ORP measurements Main Features Two point calibration front panel calibration trimmers Hand strap on the bottom for a more secure grip Compact heavy duty water resistant instrument Ideal for applications that require a custom calibration point Probe Features HI1230B pH Durable polyetherimide PEI resinbodysuitableforawiderang...

Страница 3: ...tely 150 hours of continuous use Environment 0 to 50 C 32 to 122 F 100 RH Auto shut off After 8 minutes of non use Dimensions 145 x 80 x 36 mm 5 7 x 3 1 x 1 4 Weight 230 g 8 1 oz The range may be limited by the probe s limits HI1230B pH Probe Reference double Ag AgCl Junction ceramic single Electrolyte gel Recommended Operating Temp 5 to 70 C 23 to 158 F Max Pressure 2 bar Range 0 to 12 pH Tip Sha...

Страница 4: ...set pH calibration 9 SLOPE trimmer adjusts slope pH calibration 10 CAL key enters pH calibration 11 On Off key 12 LCD display 13 Measurement mode mV or pH 14 RCA connector temperature probe General Operations Battery Replacement 1 Turn off the instrument 2 Remove the three screws on the back of the instrument to open the battery compartment 3 Remove the old battery Insert one new 9 V Alkaline batt...

Страница 5: ...s Preparation One or two point calibration can be performed using one of the following standard buffer solutions pH 4 01 6 86 NIST 7 01 9 18 NIST or pH 10 01 When a two point calibration is required use pH 7 01 or pH 6 86 buffer as first calibration point Use pH 7 01 HI7007 or pH 6 86 NIST equivalent for neutral samples pH 4 01 HI7004 for acidic samples pH 10 01 HI7010 or pH 9 18 NIST equivalent f...

Страница 6: ...meter on 2 Afterinitialization batterystatusisdisplayed Replace the battery if BATT tag is displayed blinking 3 Connect the probe pH or ORP to the meter 4 For accurate pH measurements calibrate the meter for pH before use 5 Place the tip of the electrodes 4 cm 1 1 2 into the sample 6 Press the designated key to select corresponding measurement mode i e pH for pH measurements mV for mV or ORP measu...

Страница 7: ...ering info Product description HI1230B Plastic body double junction gel combination pH electrode with BNC connector and 1 m 3 28 ft cable HI3131B Glass body refillable combination platinum ORP electrode with BNC connector and 1 m 3 28 ft cable HI7662 Temperature probe with 1 m 3 28 ft cable HI7004M pH 4 01 buffer solution 230 mL HI7006M pH 6 86 buffer solution 250 ml HI7007M pH 7 01 buffer solutio...

Страница 8: ...onments Warranty HI83141 1 is warranted for a period of two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions Electrodes and probes are warranted for a period of six months This warranty is limited to repair or replacement free of charge Damage due to accidents misuse tampering or lack of prescribed maintenance is not cov...

Отзывы: