Instruction Manual
HI 991001
Extended Range Waterproof
pH / Temperature Meter
Dear Customer,
Thank you for choosing a Hanna Instruments
Product.
This instrument is in compliance with the
directives.
All Hanna Instruments
meters are warranted
for two years
against defects in workmanship
and materials when used for their intended pur-
pose and maintained according to instructions.
The probes are warranted for a period of
six months.
This warranty is limited to repair or replacement
free of charge
.
Damage due to accidents, mis-
use, tampering or lack of prescribed maintenance
are not covered.
If service is required, contact the dealer from
whom you purchased the instrument. If under
warranty, report the model number, date of pur-
chase, serial number and the nature of the prob-
lem. If the repair is not covered by the warranty,
you will be notified of the charges incurred. If the
instrument is to be returned to Hanna Instru-
ments, first obtain a Returned Goods Authoriza-
tion Number from the Customer Service depart-
ment and then send it with shipment costs pre-
paid. When shipping any instrument, make sure it
is properly packaged for complete protection.
All rights are reserved. Reproduction in whole or
in part is prohibited without the written consent of
the copyright owner.
WARRANTY
PRELIMINARY EXAMINATION
Remove the instrument from the packing
material and examine it carefully to make
sure that no damage has occurred during
shipment. If noticeable damage is evident,
notify your dealer.
Note:
Save all packing material until you
are sure that the instrument func-
tions correctly. All defective items
must be returned in the original
packing together with the supplied
accessories.
HI 991001
is a portable, microprocessor-
based pH and temperature meter.
Main features include: extended pH and
temperature ranges; waterproof and com-
pact casing; large dual-line display; low
battery detection; automatic pH calibration
at one or two points with two memorized
buffer sets (standard or NIST); selectable
temperature unit (°C or °F).
The pH electrode has a built-in tempera-
ture sensor for simultaneous pH and tem-
perature readings, and also contains a mini
amplifier to render measurements imper-
vious to noise and electrical interferences.
Each meter is supplied complete with
•
HI 1296D
pH/temperature probe with DIN
connector and 1 m (3.3’) cable
•
calibration & cleaning solution sachets
•
3x1.5V AA alkaline batteries
•
instruction manual
•
rugged carrying case.
GENERAL DESCRIPTION
SPECIFICATIONS
Range (*)
-2.00 to 16.00 pH
-5.0 to 105.0°C / 23.0 to 221.0°F
Resolution
0.01 pH
0.1°C / 0.1°F
Accuracy (@20°C)
±0.02 pH
±0.5°C up to 60°C; ±1°C outside
±1°F up to 140°F; ±2°F outside
Typical EMC
±0.02 pH
Deviation
±0.2°C or ±0.4°F
Temperature
Automatic
Compensation
pH Calibration
Automatic, 1 or 2 point
with 2 sets of memorized buffers
(pH 4.01/7.01/10.01 or 4.01/6.86/9.18)
Probe
(included)
HI 1296D
amplified pH/temperature
probe with DIN connector and 1 m (3.3’) cable
Battery Type
3 x 1.5V AA / IEC LR6
Battery Life
Approx. 2000 hours
Auto-off
After 8 minutes of non-use
Environment
0 to 50°C (32 to 122°F);
RH max 100%
Dimensions
150 x 80 x 36 mm (5.9 x 3.1 x 1.4”)
Weight
210 g (7.4 oz.)
(*)
the temperature range is limited to 80°C (176°F) if
using the HI1296D probe.
To clean the meter, use water only.
HI 1296D
Combination amplified pH/
temperature probe with DIN
connector and 1 m (3.3’) cable
HI 7004L
pH 4.01 buffer solution, 500 mL
HI 7006L
pH 6.86 buffer solution, 500 mL
HI 7007L
pH 7.01 buffer solution, 500 mL
HI 7009L
pH 9.18 buffer solution, 500 mL
HI 7010L
pH 10.01 solution, 500 mL
HI 70300L
Storage solution, 500 mL
HI 7061L
Cleaning solution, 500 mL
HI 710007
Blue shockproof rubber boot
HI 710008
Orange shockproof rubber boot
HI 721312
Rugged carrying case
HI 76405
Electrode holder
ACCESSORIES
LCD DESCRIPTION
1. Automatic Temperature Compensation
indicator
2. Stability indicator
3. Battery percentage
4. pH calibration indicator
5. Low battery indicator
6. Selectable temperature unit
7. Secondary display
8. Primary display
9. Measuring unit for primary display
IST991001R2 07/05
w w w . h a n n a i n s t . c o m
Hanna Instruments reserves the right to
modify the design, construction and appear-
ance of its products without advance notice.