Hanna Instruments HI 98402 Скачать руководство пользователя страница 6

6

OPERATIONAL GUIDE

INITIAL PREPARATION

The instrument is supplied complete with batteries. For placing the
batteries inside the meter, see page 14.
To prepare the instrument for use, connect the Fluoride and the
reference electrodes to the input sockets on the top of the instrument.
Connect the temperature probe to the temperature socket too. The
temperature probe is used in conjunction with the F

+

 electrode for F

+

temperature compensation, but it can also be used independently to
take temperature measurements. If the probe is disconnected, 25.0
°C (72.0 °F) is displayed with °C or °F symbol blinking.
Turn the instrument ON by pressing 

On/Off

.

At start-up the display will show all
the used segments for a few seconds (or
while the button is held), followed
by the percentage indication of the
remaining battery life. The meter is
now ready to operate.
After measurement switch the instrument off, and remove the electrodes.
Rinse the 

HI 4010

 electrode with deionized water, dry it with a non

abrasive cloth, and store it dry. The 

HI 5313

 reference electrode should

be kept wet. After use, rinse it with deionized water, and store it with
a few drops of 

HI 7082

 solution in the protective cap.

The auto-off feature turns the instrument off after a set period
(default 20 min) to save battery life. To set another period or to
disable this feature, see SETUP menu on page 11.

MEASUREMENT OF FLUORIDE ION CONCENTRATION

Ensure the meter is calibrated prior to proceeding any further. In
order to obtain accurate results, calibrate the meter at a temperature
no more than ±3°C away from the sample to be measured
subsequently.

Notes

: If the fluoride electrode is new, or it has not been used for a

few days, it must be soaked in a solution of 

HI 4010-10

 (10

ppm fluoride so TISAB) for at least 3 hours

 for at least 3 hours

 for at least 3 hours

 for at least 3 hours

 for at least 3 hours

prior to measurement.

prior to measurement.

prior to measurement.

prior to measurement.

prior to measurement.

If the instrument was not calibrated (at least 1 point) or the
calibration was cleared, the LCD will display 

“----”

on the first

LCD line.

Содержание HI 98402

Страница 1: ...1 www hannainst com Instruction Manual HI 98402 Accurate Fluoride and Temperature Meter...

Страница 2: ...ot hesitate to e mail us at tech hannainst com HI 98402 is guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instru...

Страница 3: ...rolled for added precision and simplicity The calibration can be performed at one or two points The calibration points can be chosen among 1 mg L ppm 2 mg L 10 mg L 100 mg L and 1000 mg L HI 98402 is...

Страница 4: ...y holder cap 5 Liquid Crystal Display LCD 6 On Off key to turn the instrument ON and OFF 7 CAL key to enter exit calibration mode 8 Select key to select setup item STD key to select calibration standa...

Страница 5: ...s n e p m o c e r u t a r e p m e T e b o r p e r u t a r e p m e t h t i w C 5 5 o t 5 c i t a m o t u A s e d o r t c e l E 0 1 0 4 I H e l b a c 3 3 m 1 d n a C N B h t i w e d o r t c e l e e d i...

Страница 6: ...deionized water dry it with a non abrasive cloth and store it dry The HI 5313 reference electrode should be kept wet After use rinse it with deionized water and store it with a few drops of HI 7082 so...

Страница 7: ...temperature will be displayed on the upper and lower displays respectively Notes The reading is automatic temperature compensated in a 5 0 to 55 0 C temperature range The ATC tag appears on the LCD t...

Страница 8: ...d sample Wait for the reading to stabilize The temperature of the solution will be shown on the lower part of the display Notes If the temperature is out of range the last range value will be displaye...

Страница 9: ...probe to their respective connectors on top of the meter Switch the meter on by pressing the ON OFF button Note For best results it is recommended to prepare two beakers for each calibration point th...

Страница 10: ...ary LCD will display the second expected standard value together with CAL and Cal Point 2 tags After the first calibration point is confirmed submerse the Fluoride the reference electrodes and the tem...

Страница 11: ...any time The following table lists the SETUP parameters their valid values range and the factory settings default Notes Press and hold Setup then Select before CFM tag appears to toggle between standa...

Страница 12: ...nd hold down the CAL Setup keys then power on the instrument The CAL tag will appear and the secondary LCD will show 0 0 C The primary LCD will display the measured temperature or the message if the m...

Страница 13: ...Press CFM to confirm The secondary LCD will display 50 0 C Submerse the temperature probe into the sec ond vessel as close as possible to the reference thermometer Allow a few seconds for the probe to...

Страница 14: ...s follow the next steps Turn the instrument OFF Open the battery compartment cap located on the top of the instrument Remove old batteries Insert three new 1 5V AAA batteries in the battery compart me...

Страница 15: ...ing active mode and blink for warning the user MEASURE on Instrument in measurement mode SETUP on SETUP menu mode has been entered CALIBRATION on calibration mode has been entered Indication of temper...

Страница 16: ...electrode must be stored dry to prolong its life After use rinse it with deionized water dry it attentively with a non abrasive cloth and store it dry The HI 5313 reference electrode should be kept we...

Страница 17: ...Reading fluctuates could be due to dirty or blocked reference electrode junction Follow the cleaning procedure above Reading drifts submerge the tip of the reference electrode in HI 7082 solution at a...

Страница 18: ...p m e t n e k o r B e b o r p e r u t a r e p m e t e c a l p e R e b o r p e t a r b i l a c o t s l i a f r e t e M s g n i d a e r y t l u a f s e v i g r o r o E S I n e k o r B e d o r t c e l e...

Страница 19: ...eral cleaning solution 230 mL FDA bottle HI 7061L General cleaning solution 500 mL bottle HI 8061L General cleaning solution 500 mL FDA bottle HI 7073M Protein cleaning solution 230 mL bottle HI 8073M...

Страница 20: ...Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401 765 7575 E mail tech hannainst com www hannainst co...

Отзывы: