13

12

TO SCAN GLP DATA

CODE 10 - Viewing pH GLP

• Select the code 10
• The " pH GLP" message will scroll twice on LCD.
• Electrode calibration data are verified to be within acceptable

values. If not the message "Old probe" or "Dead probe" will scroll
once on LCD, depending on calibration data.

• The LCD will then display the instrument identification (ID) code.
• Press   to scan remaining data, in the following order:

last calibration date (DD.MM)
last calibration year
last calibration time (hh.mm)
electrode offset value in mV ("OF" appears in the lower LCD)
electrode slope in mV/pH ("SL" appears in the lower LCD)
first point calibration buffer
second point calibration buffer (only if a 2-point calibration has
been performed).

Note:

Data can be viewed in reverse order pressing the   key.

• Press ALT and FNC to return to function selection mode.

Code 11 - Viewing mV GLP (HI98240 only)

• Select code 11. The message "Volt GLP" will scroll twice across LCD.
• The LCD will then display the instrument identification (ID) code.
• Press   to scan remaining data, in the following order:

last calibration date (DD.MM)
last calibration year
last calibration time (hh.mm)
first calibration point
second calibration point
third calibration point (if present).

• Press ALT and FNC to return to setup mode.

CODE 12 - Printing GLP Data

• Select code 12. The message "GLP data printing" will scroll across

LCD.

• The meter will then print a complete set of the following GLP

data:

a - Instrument identification (ID) code

printed at that moment. If printout is stopped the LCD will show
the last printed sample number. It is then possible to select
another sample.

• Press the ALT and FNC keys to return to setup mode.

Viewing Logged Data

• Press RANGE to view data of the selected sample. Data will be

displayed in the following order:

pH or mV (mV reading available in 

HI98240

 only)

temperature
date
time

• If RANGE is pressed when the time is displayed, the LCD will pass

to the next Sample number.

• It is then possible to scroll the data of the next sample by

pressing RANGE or select a different sample by the arrow keys.

• to return to setup mode, press ALT and FNC when the meter

displays the sample number.

CODE 01 - Lot Summary Printout

• Select the code 01.
• The message "Lot table Printing" will scroll twice on LCD.
• The meter will then print a complete set of information based on

the data stored in memory:

a - lot number
b - logging interval
c - starting date
d - number of samples
e - starting time
f

- number and measure unit of

the logged channels.

Note:

For lot 00 the number of samples only will be printed.

DELETE SET DATA

• Select code 05 to delete the Log on Demand data or code 06 to

delete the Timed Log Data.

• A scrolling message will be displayed.
• Press the ALT and CFM keys to confirm deletion.
• It is also possible to escape without data deletion pressing the ALT

and FNC keys.

Содержание HI 98240

Страница 1: ...Instruction Manual HI 98230 HI 98240 Printing and Logging pH Meters with GLP and Smart Electrodes w w w h a n n a i n s t c o m...

Страница 2: ...meters are also equipped with a stability indicator and backlight feature for comfortable reading even in dark environments The GLP features provide a guarantee of data consistency Measurements can b...

Страница 3: ...the LOBAT indication is displayed to warn the user of this condition However the meter continues to measure correctly even when the low battery indication is displayed When the batteries become too we...

Страница 4: ...lectrode remove the pH electrode from the sample before turning the meter off If the meter is OFF detach the electrode from the meter before immersing the electrode in the storage solution ON OFF SPEC...

Страница 5: ...iption of the specific setup items their valid values and the factory settings default Code Valid values Default 00 Lot data printing scrolling 00 to 16 00 01 Print lots data summary 02 Printer enable...

Страница 6: ...sample is reached press and hold the ALT and PAPER keys until the printer stops During printout the LCD will display the sample number being Code Valid values Default 10 Show pH GLP data 11 Show ORP G...

Страница 7: ...age GLP data printing will scroll across LCD The meter will then print a complete set of the following GLP data a Instrument identification ID code printed at that moment If printout is stopped the LC...

Страница 8: ...ferent samples it is recommended to rinse the electrode thoroughly with deionized water or if not available tap water first and then with some of the next sample to condition the electrode before imme...

Страница 9: ...ORP electrode to the meter and press ON OFF to power on the instrument The meter automatically sets the mV range If the ORP electrode is not connected the LCD will display the no probe message and th...

Страница 10: ...ng will be printed beside the date Note A dashed line indicates that the calibration has never been performed 6 If password is correct or not set the meter will display 7 01pH on the up per LCD and b1...

Страница 11: ...when LCD is displaying mV 3 User is prompted to enter password if it has been set to a value different from 0000 otherwise skip to step 6 4 Enter the password with the arrow keys 5 Press ALT and CFM t...

Страница 12: ...ly enabled in order to return to logging with printing mode LOG ON DEMAND In measuring or Timed log mode press LOG to store the current reading The LCD will display Two different modes to print log da...

Страница 13: ...the meter on again CALIBRATION ALARM TIME OUT The calibration alarm time out is available only for pH calibration It is possible to set through setup code 13 the number of days before the next requir...

Страница 14: ...t is possible to make any elaboration available with software e g graphics statistic analysis HI 92000 offers a variety of features and is provided with an on line help to support the user throughout...

Страница 15: ...lot was the 16th the new logging period restarts from lot 01 overwriting previously logged data When Timed logging memory is full the meter overwrites the oldest lot data progressively reducing the o...

Страница 16: ...I 98240 can also be used with other adapters In this case remember to check the correct polarity of your adapter before connecting it to the meter BATTERY REPLACEMENT TEMPERATURE RESISTANCE CORRELATIO...

Страница 17: ...own could be due to Clogged Dirty Junction refer to the Cleaning Procedure above Loss of shielding due to low electrolyte level in refillable electrodes only refill with fresh HI 7071 for single junct...

Страница 18: ...7 01 buffer solution 500 mL FDA bottle HI 8009L pH 9 18 buffer solution 500 mL FDA bottle HI 7009L pH 9 18 buffer solution 500 mL bottle HI 7010L pH 10 01 buffer solution 500 mL bottle HI 8010L pH 10...

Страница 19: ...10005 Voltage adapter 115 Vac 12 Vdc HI 710006 Voltage adapter 230 Vac 12 Vdc HI 710031 Rugged carrying case HI 710034 Paper roll 10 pcs HI 710035 Ink cartridge HI 76405 Electrode holder HI 8427 pH mV...

Страница 20: ...s in workmanship and materials when used for their intended purpose and maintained according to instructions The electrodes and the probes are warranted for a period of six months This warranty is lim...

Страница 21: ...42 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 0...

Отзывы: