Meter Calibration

From measurement mode, press 
and hold the ON/OFF button 
until “CAL” is displayed.

When “7.01” blinks on the display, 
place the tip of the probe into a  
pH 4.01, 7.01, or 10.01 buffer 
solution.

For one-point calibration using pH 4.01 or pH 10.01 

buffer go to procedure B 

For one or two-point calibration using pH 7.01 

buffer go to procedure A

A

B

When the reading is stable, 
the stability indicator will 
disappear.“Sto” will be displayed 
when the calibration is saved.

If pH 4.01 or pH 10.01 buffer solution 
is used as the first point

 the value of 

the buffer is recognized and displayed 
with the blinking stability indicator.

Meter will exit to measurement 
mode and the calibration tag will 
be displayed.

One-Point Calibration with pH 4.01 or pH 10.01

B

On

e-

Po

in

t

Meter will exit to measurement 
mode and the calibration 
tag will be displayed.

pH 4.01 will then blink 
on the display.

Tw

o-

Po

in

t

Use pH 4.01 or pH 10.01 to 
perform a two-point calibration. 
The value is automatically 
recognized and displayed with 
the blinking stability indicator.

When the reading is stable, the 
stability indicator will disappear. 
“Sto” will be displayed when 
the calibration is saved.

Meter will exit to measurement 
mode and the calibration 
tags will be displayed.

A

One or Two-Point Calibration with pH 7.01

If pH 7.01 buffer solution is used as 
the first point

 the buffer is recognized 

with the blinking stability indicator.

When the reading is stable, the 
stability indicator will disappear 
and pH 7.01 will be calibrated. 

If pH 7.01 is the only calibration 
point,

 finish one-point 

procedure at right. 

If using pH 4.01 or pH 10.01 
as a second point,

 continue 

two-point procedure at right.

Warranty

The meter is warranted for a period of one year against defects in 
workmanship and materials when used for their intended purpose and 
maintained according to instructions. The electrode is warranted for a 
period of six months. This warranty is limited to repair or replacement free 
of charge. Damage due to accidents, misuse, tampering or lack of prescribed 
maintenance is not covered. If service is required, contact your local Hanna 
Instruments Office. If under warranty, report the model number, date of 
purchase, serial number and the nature of the problem. If the repair is 
not covered by the warranty, you will be notified of the charges incurred. 
If the instrument is to be returned to Hanna Instruments, first obtain a 
Returned Goods Authorization (RGA) number from the Technical Service 
department and then send it with shipping costs prepaid. When shipping 
any instrument, make sure it is properly packaged for complete protection.

Before using Hanna Instruments products, make sure that they are entirely 
suitable for your specific application and for the environment in which they 
are used. Any variation introduced by the user to the supplied equipment 
may degrade the meter's performance. Avoid touching the electrode area.
To avoid damages or burns, do not put the meter in microwave oven. For 
yours and the meter's safety do not use or store the meter in hazardous 
environments.

Recommendations for Users

All rights are reserved. Reproduction in whole or in part is prohibited 

without the written consent of the copyright owner, Hanna Instruments 

Inc., Woonsocket, Rhode Island, 02895, USA.

To obtain the highest accuracy for measurements it is important to follow 
these tips:

•  Calibration is only as good as the buffer being used. The pH buffer values 

change over time once the sachets are opened. Fresh buffer should be 
used for each calibration.

•  The probe should be rinsed with purified water each time before placing in 

buffer or sample to be tested.

•  When the meter is not in use it is important to add several drops of storage 

solution to the protective cap to keep the probe hydrated. If storage 
solution is not available, pH 4.01 or pH 7.01 buffer can be used.

•  For improved accuracy it is recommended to calibrate to a minimum of two 

points. It is important to use buffers that bracket the expected value of the 
sample to be tested. For example, if the expected value is pH 8, the meter 
should be calibrated using pH 7.01 and pH 10.01 buffers.

•  It is important to calibrate and measure samples at the same temperature. 

A dramatic change in temperature between buffer solutions and samples 
to be tested will give inaccurate readings

.

Care and Maintenance

IST98103N  06/18

To change the CR2032 Li-ion battery, turn the battery cover located on the 
back of the meter counterclockwise to unlock. Remove cover and replace 
with + side facing up.

Note: Batteries should only be replaced in a safe area using the battery 
type specified in this instruction manual. Old batteries should be 
disposed in accordance with local regulations.

Battery Replacement

Electrode Storage Solution

Code

Description

HI70300L

electrode storage solution, 500 mL bottle

HI70300M

electrode storage solution, 230 mL bottle

Electrode Cleaning Solution

Code

Description

HI700601P

general purpose cleaning solution,  
20 mL sachets (25 pcs.)

Electrode

Code

Description

HI1271

pH electrode for Checker

pH Buffer Solution

Code

Description

HI70004P

pH 4.01 buffer solution, 20 mL sachets (25 pcs.)

HI70007P

pH 7.01 buffer solution, 20 mL sachets (25 pcs.)

HI70010P

pH 10.01 buffer solution, 20 mL sachets (25 pcs.)

HI77400P

pH 4.01 & 7.01 buffer solution, 20 mL sachets 
(10 pcs., 5 ea.)

HI770710P

pH 10.01 & 7.01 buffer solution, 20 mL sachets 
(10 pcs., 5 ea.)

Accessories

Содержание HI 98103

Страница 1: ...HI98103 pH Tester INSTRUCTION MANUAL ...

Страница 2: ...re that the instrument functions correctly All defective items must be returned in the original packaging with the supplied accessories Preparation ThepHelectrodeisshippeddry BeforeusingtheChecker removetheprotective cap and condition the electrode by soaking the tip bottom 4 cm 1 5 in pH 7 01 buffer solution for several hours Then follow the calibration procedure Do not be alarmed if white crysta...

Страница 3: ...o off 8 minutes 60 minutes or can be disabled Environment 0 to 50 C 32 to 122 F RH 95 max Dimensions 50 x 174 x 21 mm 2 x 6 8 x 0 9 Weight 50 g 1 8 oz Liquid Crystal Display ON OFF Button pH Probe Maximum Immersion Level Recommended Immersion Level 4 cm 1 5 Operation Press the ON OFF button to turn the meter on All tags will be displayed The meter will go into measurement mode current reading and ...

Страница 4: ...en in the correct fresh buffer solution then the probe should be cleaned Place the probe in the HI700601 cleaning solution for 15 minutes Rinse with purified water and place in storage solution for 1 hour before calibrating If the Err message persists then the HI1271 probe should be replaced Clear Calibration Err Message Battery Indicator From measurement mode press and hold the ON OFF button The ...

Страница 5: ... pH 4 01 or pH 10 01 B One Point Meter will exit to measurement mode and the calibration tag will be displayed pH 4 01 will then blink on the display Two Point Use pH 4 01 or pH 10 01 to perform a two point calibration The value is automatically recognized and displayed with the blinking stability indicator When the reading is stable the stability indicator will disappear Sto will be displayed whe...

Страница 6: ...meter is not in use it is important to add several drops of storage solution to the protective cap to keep the probe hydrated If storage solution is not available pH 4 01 or pH 7 01 buffer can be used Forimprovedaccuracyitisrecommendedtocalibratetoaminimumoftwo points It is important to use buffers that bracket the expected value of the sample to be tested For example if the expected value is pH 8...

Страница 7: ...department and then send it with shipping costs prepaid When shipping any instrument make sure it is properly packaged for complete protection All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA IST98103N 06 18 Electrode Storage Solution Code Description HI70300L electrode ...