Hanna Instruments HI 96715C Скачать руководство пользователя страница 8

14

15

• Press 

ZERO/CFM

 and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• After a few seconds, the display will show

“-0.0-”

. The meter is now zeroed and

ready for measurement.

• Remove the cuvette, open it and add 4

drops of 

HI 93715A-0

 First Reagent (6

drops in case of seawater analysis). Swirl
the cuvette gently to mix.

• Add 4 drops of 

HI 93715B-0

 Second

Reagent to the cuvette (10 drops in case
of seawater analysis). Replace the cap
and swirl the cuvette to mix.

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/

READ/

READ/

READ/

READ/ /TIMER

/TIMER

/TIMER

/TIMER

/TIMER 

for three

seconds.

 The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.

Alternatively, wait for 3 minutes and 30
seconds and just press READ/

READ/

READ/

READ/

READ/ /TIMER

/TIMER

/TIMER

/TIMER

/TIMER

.

In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

• At the end of measurement, the instrument

directly displays concentration in mg/L of
ammonia nitrogen (NH

3

-N) on the LCD.

• To convert the reading to mg/L of

ammonia (NH

3

), multiply by a factor of

1.216.

VALIDATION PROCEDURE

Use the validation procedure to ensure that the instrument is properly calibrated.

Warning:

 Do not validate the instrument with any standard solutions other than the HANNA 

CAL

CHECK

TM

 Standards, otherwise erroneous results will be obtained. For accurate validation, please

perform test at room temperature: 18 to 25°C (64.5 to 77.0°F)

• Turn the meter on by pressing 

ON/OFF

.

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.

INTERFERENCES

Organic compounds like: chloramines, various aliphatic and aromatic amines, glycine,
(positive error).
To eliminate these interferences distillation is required.

Organic compounds like: aldehydes, alcohols (e.g. ethanol) or acetone (negative error).
To eliminate these interferences distillation is required.

Sulfide (S

2-

): may cause turbidity.

Hardness above 1 g/L as Calcium Carbonate (CaCO

3

).

Содержание HI 96715C

Страница 1: ...1 Instruction Manual HI 96715C Ammonia Medium Range ISM 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Страница 2: ...t carefully before using the meter All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA PRELIMINARY EXAMINATION Please examine this product carefully Make sure that the instrument is not damaged If any damage occurred during shipment please notify your Dealer Each HI 96715 I...

Страница 3: ... an adaptation from ASTM Manual of Water and Environmental Technology D1426 93 Nessler Method The reagents are liquid form and are supplied in bottles The amount of reagent is precisely dosed to ensure the maximum repeatability degree Celsius degree Fahrenheit milligrams per liter mg L is equivalent to ppm parts per million milliliter Light Emitting Diode Liquid Crystal Display American Society fo...

Страница 4: ...er is based on a Light Emitting Diode LED combined with a narrow band interference filter to guarantee both high performance and reliable results HI 96715 block diagram optical layout A microprocessor controlled LED emits radiation which is first optically conditioned and beamed to the sample contained in the cuvette The optical path is fixed by the diameter of the cuvette Then the light is spectr...

Страница 5: ...ion date or restore factory calibration DISPLAY ELEMENTS DESCRIPTION 1 The measuring scheme lamp cuvette detector appears during different phases of zero or reading measurement 2 Error messages and warnings 3 The battery icon shows the charging level of the battery 4 The hourglass appears when an internal checking is in progress 5 Status messages 6 The chronometer appears when the reaction timer i...

Страница 6: ...g 0 00 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvette for reference zero and measurement Over Range A flashing value of the maximum concentration indicates an over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test c during calibration procedure Standar...

Страница 7: ...on times reported in this manual are referred to 20 C 68 F As a general rule of thumb they should be doubled at 10 C 50 F and halved at 30 C 86 F In order to maximize accuracy prior to a measurement follow the validation procedure to be sure that the instrument is properly calibrated If necessary calibrate the instrument MEASUREMENT PROCEDURE STARTUP Prepare the instrument for measurement as follo...

Страница 8: ...oth cases the lamp cuvette and detector icons will appear on the display depending on the measurement phase At the end of measurement the instrument directly displays concentration in mg L of ammonia nitrogen NH3 N on the LCD To convert the reading to mg L of ammonia NH3 multiply by a factor of 1 216 VALIDATION PROCEDURE Use the validation procedure to ensure that the instrument is properly calibr...

Страница 9: ...ications as reported in the CAL CHECKTM Standard Certificate If the value is found out of the specifications please check that the cuvettes are free of fingerprints oil or dirt and repeat validation If results are still found out of specifications then recalibrate the instrument Turn the meter on by pressing ON OFF When the beeper sounds briefly and the LCD displays dashes the meter is ready Press...

Страница 10: ...w seconds the display will show 0 0 The meter is now zeroed and ready for calibration The blinking READ asks for reading calibration standard Remove the cuvette Place the CAL CHECKTM Standard HI 96715 11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove Press READ TIMER and the lamp cuvette and detector icons will appear on the display de...

Страница 11: ...ery indicator with three levels as follows 3 lines for 100 capacity 2 lines for 66 capacity 1 line for 33 capacity Empty battery icon blinking if the capacity is under 10 If the battery is empty and accurate measurements can t be taken anymore the instrument shows dead batt dead batt dead batt dead batt dead batt and turns off To restart the instrument the battery must be replaced with a fresh one...

Страница 12: ...e instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs prepaid When shipping any instrument make sure it is properly packaged for complete protection To validate your warranty fill out and return the enclosed warranty card within 14 days from the date of purchase Hanna Instrume...

Отзывы: