
19
• Alternatively, wait for 45 seconds then
just press
READ/ /TIMER.
In both
cases, the lamp, cuvette and detector
icons will appear on the display,
depending on the measurement phase.
• At the end of measurement, the
instrument directly displays concentration
in mg/L of cyanuric acid on the LCD.
VALIDATION PROCEDURE
Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:
Do not validate the instrument with any standard solutions other than the HANNA
CAL
CHECK
TM
Standards, otherwise erroneous results will be obtained.
Note
: The validation is performed only for the selected parameter. For full validation of the
instrument, the following procedure must be performed for each parameter.
• Turn the meter on by pressing
ON/OFF
.
• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
• Place the
CAL CHECK
TM
Standard
HI 96710-11
Cuvette A into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
• Press
ZERO/CFM
and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
• After a few seconds, the display will
show
“-0.0-”
. The meter is now zeroed
and ready for validation.
• Remove the cuvette.