FUNCTIONAL DESCRIPTION

1) Primary LCD:

When the meter is switched on, the LCD
shows all segments for a few seconds.

It then displays four dashes to indicate
“ready to measure”.

In “Read” and “Zero” mode, “SIP” is shown
to indicate “Sample In Progress”.

The upper level also indicates the concen-
tration or turbidity value of the sample, as
well as different diagnostic modes, such
as “-BA-” for low battery.

A substance absorbs a color complimentary
to the color it emits. For example, a
substance appears yellow because it absorbs
blue light. As a result, the Hanna meters use
LED’s with specific wavelengths to measure
samples.

The optical distance (d) is measured by the
internal diameter of the cuvet containing the
sample.

The photoelectric cell collects the radiation 

I

that is not absorbed by the sample and
converts it into an electric current.

The microprocessor converts the value into
the desired measuring unit and displays it on
the LCD.

The measurement process is done in two
phases: setting the meter to zero and actual
measurement.

The cuvet is an optical element and hence
has an important role in the measurement
process. Both the measurement and the
calibration cuvets must be optically identical
to provide the same measurement conditions.

It is also important that the surface of the
cuvet is clean and free from scratches or
dents, in order to avoid measurement
interference due to unwanted reflection and
absorption of light.

It is recommended that wherever possible the
cuvet walls are not touched by the operator.

Furthermore, in order to maintain the same
conditions during the zeroing and the
measuring phases, it is necessary to close
the cuvets to prevent any contamination.

Содержание HI 93114

Страница 1: ...Instruction Manual HI 93114 Turbidity Chlorine Measurements w w w h a n n a i n s t c o m...

Страница 2: ...ectly Any defective item must be re turned in its original packaging with the supplied accessories 1 NTU Nephelometric Turbidity Unit 1 FTU Formazine Turbidity Unit DearCustomer ThankyouforchoosingaHA...

Страница 3: ...PA compliant turbidity mea surements are performed quickly and reliably withthislab grademicroprocessor basedfield instrument HI93114isparticularlyusefulforpotablewater measurementsandmunicipalities W...

Страница 4: ...nce determined by the emitted radiation I An LED Light Emitting Diode emits radiation at a relatively narrow spectrum supplying the system with the intensity Io LED EMITTED LIGHT CUVET MICROPROCESSOR...

Страница 5: ...the radiation I that is not absorbed by the sample and converts it into an electric current The microprocessor converts the value into the desired measuring unit and displays it on the LCD The measure...

Страница 6: ...ored data date time value 13 ALT key Activates alternative functions 14 key Scrolls downwards through the parameters to be measured 2 Secondary LCD The three digit lower level shows the cur rent measu...

Страница 7: ...ur minute data will be stored in memory SPECIFICATIONS 1 NTU 1 FTU In calibration and diagnostic modes decreases the blink ing digit by one If pressed together with ALT while the meter is in logging m...

Страница 8: ...e notch on the cap is positioned securely into the groove Press ZERO CURRENT TIME DATE RECALL To recall current TIME DATE press and hold the ALT and READ keys together The cur rent time and an intermi...

Страница 9: ...e been stored Add the appropriate reagent into the blank sample cuvet Shake and allow a few sec onds for color to develop For the Free or Total Chlorine add the con tent of its respective packet Repla...

Страница 10: ...Turn the meter on and press both ALT and CAL momentarily The LCD will show four dashes on the upper and d 00 on the lower part of the LCD with the second 0 blinking Using the and keys change the lowe...

Страница 11: ...vet holder and make sure that the notch is positioned securely into the groove Press the key The display will indicate SIP again The display will blink Lt for several sec onds indicating that the LED...

Страница 12: ...point customized calibration To enter the colorimetric calibration mode press momentarily both the ALT and CAL keys The LCD will show four dashes and d00 After the second calibration point is memo riz...

Страница 13: ...ion point is memo rized the unit will store the calibration data time and date in the EEPROM while intermit tently indicating CAL and Stor for several seconds Using the and keys set the lower part of...

Страница 14: ...brates span in turbidimetric mode 31 Calibrates span in colorimetric mode 40 Clears the logged memory The following diagnostic modes are reserved for authorized service technicians 00 Shows the Blank...

Страница 15: ...r can easily turn the log ging mode on and off review the logged memory review the current lot number and clean the buffer memory HI 93114 also reminds the user if its memory is full TURNING THE LOG M...

Страница 16: ...ngs in its non volatile memory and the display will flash CAL and Stor alternately for several sec onds USER SELECTABLE SHUTDOWN With HI 93114 the users can customize the shutdown time to save power T...

Страница 17: ...er ies as soon as possible BA Exhausted batteries change all bat teries immediately LO Low light level is received during the zeroing procedure check the cuvet for scratches and ensure that the sample...

Страница 18: ...according to instructions This warranty is limited to repair or replace ment free of charge Damages due to acci dents misuse tampering or lack of prescribed maintenance are not covered If service is r...

Страница 19: ...3 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 27...

Отзывы: