START UP

• Place the instrument on a flat table. Do not place the instrument on direct sun light.
• Connect the titrator to mains socket with ground connection and the correct voltage and

frequency. See the label on the instrument rear for this.

• Place the peristaltic pump tube on the pump. See the Pump Tube Replacement section for the

procedure.

• Remove the reagent bottle cap and place the bottle cap of the tubes set. Place the reagent

bottle in the appropriate place on the titrator top.

• Connect the tubes with the peristaltic pump (inlet tube is connected with the reagent bottle,

outlet tube is connected with the dosing tip).

• Turn the instrument ON using the power switch from the rear panel of the instrument and

wait until it displays dashes.

12

The instructions listed below should be carefully followed during testing to ensure best accuracy.

• Purge the peristaltic pump to have fresh titrant when starting a new analysis or calibration.
• Calibrate the peristaltic pump before performing an analysis.
• Analyze the wine immediately after the sample is obtained.
• Clean the electrode with the 

HI 700635

 or 

HI 700636

 cleaning solution, specially made for

wine industry, if it was unused for a long time.

TIPS FOR AN ACCURATE MEASUREMENT

• Fill the 50 mL beaker up to the 50 mL mark with

the wine sample, place the stir bar into the beaker
and put the beaker in the appropriate place on the
minititrator top.

• Fill the 20 mL beaker up to the 5 mL mark with the

HI 84100-53

 Acid Reagent and add the content to

the 50 mL beaker.

• Add the content of one powder packet of 

HI 84100-54

Stabilizer into the beaker.

• Place the probe holder on the top of the beaker and

secure it by turning clockwise.

FREE SO

2

 MEASUREMENT PROCEDURE

Warning:

Make sure the pump was calibrated before performing sulphur dioxide wine sample
analysis.

Содержание HI 84100

Страница 1: ...851 9129 0 Fax 07851 9129 99 Greece Tel 210 823 5192 Fax 210 884 0210 Indonesia Tel 21 4584 2941 Fax 21 4584 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malays...

Страница 2: ...ACCESSORIES 18 WARRANTY 19 19 Hanna Instruments reserves the right to modify the design construction and appearance of its products without advance notice HI 84100 is warranted for two years against...

Страница 3: ...ffective interface for the user By simply pressing the START STOP button the instrument will automatically make the titration up to the equivalence point The result is immediately displayed in conveni...

Страница 4: ...ted by the human senses is about 2 0 ppm This is also the level which is needed for maximum protection of your wine This is particularly true in the case of sweet and most notably botrytised wines The...

Страница 5: ...PREPARATION PROCEDURE Remove the protective cap of the ORP electrode HI 3148B DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT This is normal with electrodes They will disappear when rinsed with water D...

Страница 6: ...volume has to be precisely known 15 PUMP CALIBRATION PROCEDURE Warning The calibration of the pump must be performed each time the pump tube the reagent bottle or the pH electrode is changed It is rec...

Страница 7: ...ng clockwise Immerse the ORP electrode approximately 2 cm 0 8 into the sample to be tested while paying attention to not touch the stir bar Insert the dosing tip in the appropriate holder place and pa...

Страница 8: ...e paying attention to not touch the stir bar Insert the dosing tip in the appropriate holder place and pay attention to not be immersed into solution Press the START STOP button to start the titration...

Страница 9: ...ed during testing to ensure best accuracy Purge the peristaltic pump to have fresh titrant when starting a new analysis or calibration Calibrate the peristaltic pump before performing an analysis Anal...

Страница 10: ...urging mode message PUMP CALIBRATION MESSAGES 11 This error message appears when the sample concentration exceeds 400 ppm The used standard solution is wrong This error message appears when the input...

Отзывы: