Instruction Manual
HI 83141V
Complete Portable Kit
for pH, mV & Temperature
Measurements in
Wine, Must and Grape Juice
w w w . h a n n a i n s t . c o m
WARRANTY
All Hanna Instruments
meters are warranted for two years
against
defects in workmanship and materials when used for their intended pur-
pose and maintained according to instructions.
The electrodes and the probes are warranted for a period of six
months.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed main-
tenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase,
serial number and the nature of the failure.
If the repair is not covered by the warranty, you will be notified of the
charges incurred. If the instrument is to be returned to Hanna Instruments,
first obtain a Returned Goods Authorization Number from the Customer
Service department and then send it with shipment costs prepaid.
When shipping any instrument, make sure it is properly packaged for
complete protection.
Dear Customer,
Thank you for choosing a Hanna product.
Please read carefully this instruction manual before using the
meter. If you need additional technical information, do not
hesitate to e-mail us at
.
This instrument is in compliance with the
directives.
Remove the kit from the packing material and examine it
carefully to make sure that no damage has occurred during
shipment. If there is any damage, notify your dealer.
The kit includes:
• HI 83141
portable pH meter
• HI 1048B
pH electrode
•
HI 7669AW
temperature probe
• pH 3 and pH 7 buffer solutions (20 mL sachet each)
•
HI 700635
and
HI 700636
specific cleaning solutions
• electrode refilling solution (30 mL bottle) and syringe
• 9V battery, screwdriver and instructions.
Note:
Conserve all packing material until the instrument has
been observed to function correctly. Any defective item
must be returned in its original packing.
PRELIMINARY EXAMINATION
FUNCTIONAL DESCRIPTION
SPECIFICATIONS
1) BNC connector for pH (or ORP) electrode.
2) Liquid Crystal Display.
3)
mV
key, to display the mV (ORP) readings when using
an ORP electrode or the mV equivalent to pH values
when using a pH electrode.
4)
T
rimmer
for pH OFFSET calibration.
5)
T
rimmer for pH SLOPE calibration.
6)
°C
key, to display the temperature measurement.
7)
pH
key, to display the pH value.
8)
ON/OFF
key.
9) Socket for
HI 7669AW
temperature probe.
Range
0.00 to 14.00 pH / ±1999 mV
0.0 to 100.0°C
Resolution
0.01 pH / 1 mV / 0.1°C
Accuracy
(@20°C/68°F)
±0.01 pH / ±1 mV / ±0.4°C
Typical EMC Deviation
±0.07 pH / ±5 mV / ±1°C
pH Calibration
Manual, 2 point, through
offset (±1 pH) and slope (85 to 105%) trimmers
ORP Calibration
Factory calibrated
Temperature Calibration
Factory calibrated
Temperature Compensation
Automatic, from 0 to 70°C
Probes
(
included
)
HI 1048B
pH-electrode
HI 7669AW
temperature probe
Battery Type
1 x 9V, alkaline
Battery Life
100 hours of continuous use
Environment
0 to 50°C (32 to 122°F);
RH max 95% non-condensing
Dimensions
164 x 76 x 45 mm (6.5 x 3.0 x 1.8")
Weight
180 g (6.3 oz.)
ACCESSORIES
HI 1048B
Glass body, double-junction, refillable pH elec-
trode with BNC connector and 1 m (3.3’) cable
HI 3131B
Glass-body, refillable, platinum ORP electrode
with BNC connector and 1 m (3.3’) cable
HI 7669AW
Temperature probe with 1 m (3.3’) cable
HI 5003
pH 3.00 buffer solution, 500 mL bottle
HI 5068
pH 6.86 buffer solution, 500 mL bottle
HI 5007
pH 7.01 buffer solution, 500 mL bottle
HI 70300L
Storage solution, 500 mL bottle
HI 70635L
Cleaning solution for wine deposits, 500 mL
HI 70636L
Cleaning solution for wine stains, 500 mL
HI 7082
Electrolyte solution, 3.5M KCl, 4 x 30 mL bottle
HI 7091L
Reducing pretreatment solution, 500 mL
HI 7092L
Oxidizing pretreatment solution, 500 mL
HI 731326
Calibration screwdriver (20 pcs)
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the environment in which
it is used. Operation of this instrument in residential area could cause unacceptable interference
to radio and TV equipment, requiring the operator to take all necessary steps to correct
interference.
The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid
touching this glass bulb at all times.
To maintain the EMC performance of equipment, the recommended cables noted in the
user's manual must be used. Any variation introduced by the user to the supplied equipment
may degrade the instruments' EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement surface
exceed 24 Vac or 60 Vdc.
To avoid damage or burns, do not perform any measurement in microwave ovens.
IST83141V 07/05