Instruction Manual

HI 7101 • HI 7102

RH Calibration Kits

Dear Customer,

Remove the calibration kit from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.

Note:

Conserve all packing material until the kit has been
observed to function correctly. Any defective item must
be returned in its original packing.

PRELIMINARY EXAMINATION

Thank you for choosing a Hanna product.
Please read this instruction manual carefully before using
the instrument. This manual will provide you with the
necessary information for correct operation.
If you need additional technical information, do not hesitate
to e-mail us at 

[email protected]

.

HI 7101

 and 

HI 7102

 portable

chambers allows you to calibrate
your hygrometers. These kits include
two thermally isolated chambers,
each one equipped with a cap and
three bottles containing the appro-
priate precalibrated saturated salts to produce a known RH
value: two 15-gram bottles of Lithium Chloride (LiCl) and
one 33-gram bottle of Sodium Chloride (NaCl).

HI 7101

 is used with 

HI 8564

HI 9064

HI 9065

,

HI 9161

 and 

HI 91610

.

HI 7102

 is used with 

HI 93640

 and 

HI 8666

.

They measure only 190 x 115 x 110 mm.
You can check the status of your probe

 

and perform a quick

recalibration (with an accuracy of ±5%) by using the
Hanna mini-calibration chambers 

HI 7101 

and 

HI 7102.

GENERAL DESCRIPTION

P

REPARING

  

THE

  C

ALIBRATION

  S

OLUTIONS

:

• Pour approximately 26 cc of distilled

water into a glass container.

• Immerse this container into a bath of ice

and water and shake briefly.

• Slowly add the contents of a 

HI 7111

bottle containing LiCl into the glass con-
tainer while continuing to shake.

• When the salt has dissolved completely, add all the

content of the second bottle of 

HI 7111

.

• Allow the solution obtained to cool, and

then pour it into the chamber marked
"RH11.1%", making sure that no resi-
due remains on the walls of the glass
container.

• Seal the chamber well when not in use, as the LiCl

solution is extremely hygroscopic and tends to capture
the humidity present in the air causing the solution to
expand in volume and to overflow from the container.

CALIBRATION

Li Cl

11.1%

Na Cl

75.4%

CALIBRATION KIT

Li Cl

11.1%

Na Cl

75.4%

CALIBRATION KIT

H

2

O

26 cc

Li Cl

Li Cl

11.1%

Na Cl
75.4%

CALIBRATION KIT