53
PH ELECTRODE MAINTENANCE
For refillable electrodes:
Refill the reference chamber with fresh electrolyte (
HI7071
or
HI8071
for
single junction and
HI7082
or
HI8082
for double junction electrodes).
Allow the electrode to stand upright for 1 hour.
Follow the Storage Procedure above.
pH ELECTRODE CLEANING PROCEDURE
• General
Soak in Hanna Instruments
HI7061
or
HI8061
General
Cleaning Solution for approximately ½ hour.
• Protein
Soak in Hanna Instruments
HI7073
or
HI8073
Protein
Cleaning Solution for 15 minutes.
• Inorganic
Soak in Hanna Instruments
HI7074
Inorganic Cleaning
Solution for 15 minutes.
• Oil/grease
Rinse with Hanna Instruments
HI7077
or
HI8077
Oil
and Fat Cleaning Solution.
IMPORTANT:
After performing any of the cleaning procedures, rinse the
electrode thoroughly with distilled water, refill the reference chamber
with fresh electrolyte (not necessary for gel‑filled electrodes) and soak
the electrode in
HI70300
or
HI80300
Storage Solution for at least
1 hour before taking measurements.
EC Probe Maintenance
Rinse the probe with clean water after measurements. If a more thorough
cleaning is required, remove the probe sleeve and clean the probe with a
cloth or a nonabrasive detergent. Make sure to reinsert the sleeve onto
the probe properly and in the right direction. After cleaning the probe,
recalibrate the instrument.
The platinum rings support is made of glass. Take great care while
handling the probe.