50
GLDxxxff
: Requests the records of the “xxx” lot number. The records are
sent in frames of 10 records; “ff” is the frame number (01
first frame). (Example: Lot 13 has 53 records. The records will
be sent in 6 frame, 5 with10 records and 1 with 3 records.)
Command Parameters:
xxx ‑ Lot number (eq: for lot number 1 xxx = 001)
ff ‑ requested frame number ‑ first frame is labeled 01
Errors:
• “Err3” log on demand empty.
• “Err4” requested set parameter is not available.
• “Err5” command argument is wrong.
• “Err6” requested range not available.
• “Err7” meter in log mode.
• “Err8” is sent if instrument is not in measurement mode.
• “NAK” (21) character is sent when the instrument receives
an unknown or a corrupted command.
Note:
P ‑ request for pH range.
M ‑ request for mV range.
E ‑ request for EC range.
T ‑ request for TDS range.
N ‑ request for NaCl range.