
48
• EC calibration data (if available), which contains:
• the number of calibrated standards (1 char)
• the offset factor, with sign and decimal point
(7 chars)
• the cell constant, with sign and decimal point
(7 chars)
• the calibration time,
yymmddhhmmss
(12 chars)
• standards information (for each standard)
• standard value, with sign and decimal point
(7 chars).
• buffer unit (2 chars; 00‑
µ
S; 01‑mS)
• Reference Temperature with and decimal point
(4 chars)
• Temperature Compensation mode (2 chars)
• 00 ‑ no temperature compensation
• 01 ‑ automatic temperature compensation
• 00 ‑ manual temperature compensation
• TC coefficient with sign and decimal point
(4 chars)
• calibration time,
yymmddhhmmss
(12 chars).
• Na Cl Calibration data
• the number of calibrated data (1 char)
• salinity coefficient, with sign and decimal point
(7 chars)
• Cell constant, with sign and decimal point
(2 chars)
• the calibration time (2 chars)
• buffer information, for each buffer:
• Reference Temperature with and decimal point
(4 chars)
• Temperature Compensation mode (2 chars)
• 00 ‑ no temperature compensation
• 01 ‑ automatic temperature compensation
• 00 ‑ manual temperature compensation
• TC coefficient with sign and decimal point
(4 chars)
• calibration time,
yymmddhhmmss
(12 chars).
PAR
Requests the setup parameters setting.
The answer string contains:
• Instrument ID (4 chars)