
23
GOOD LABORATORY PRACTICE (GLP)
GLP is a set of functions that allows storage and retrieval of data regarding
the maintenance and status of the system.
All data regarding pH, Rel mV, EC and NaCl calibration is stored for the
user to review when necessary.
EXPIRED CALIBRATION
For pH calibration, this instrument allows the user to set the number of
days before the next required pH calibration. This value can be set from
1 to 7 days. The default setting is OFF (disabled).
When calibration has expired the “
CAL
” and “
DUE
” tags will blink to
warn the user that the instrument should be recalibrated.
For example, if a 4 days time out has been selected, the instrument will
issue the alarm exactly 4 days after the last calibration.
If the expiration value is changed (e.g. to 5 days), then the alarm will
be immediately recalculated and appear 5 days after the last calibration.
Note: If the instrument was not calibrated, the “CAL” “DUE” tags
will be displayed even if the feature is disabled in SETUP menu.
pH CALIBRATION DATA
Calibration data is stored automatically after a successful calibration.
To view the pH calibration data, press
GLP
when the instrument is in
pH measurement mode.
The instrument will display the time (hh:mm) of the last calibration.
Use the
ARROW
keys to scroll through the calibration data:
• The date (mm.dd.yyyy).
• The pH calibration offset.