background image

INSTRUCTION MANUAL

HI98331

Direct Soil Conductivity & 

Temperature Meter with 

built-in Stainless Steel  

Conductivity Probe

Warranty

The GroLine is warranted for a period of one year 
against defects in workmanship and materials 
when used for its intended purpose and maintained 
according to instructions. This warranty is limited to 
repair or replacement free of charge. Damage due to 
accidents, misuse, tampering or lack of prescribed 
maintenance is not covered. If service is  required, 
contact your local Hanna Instruments Office. If 
under warranty, report the model number, date 
of purchase, serial number and the nature of the 
problem. If the repair is not covered by the warranty, 
you will be notified of the charge incurred. If the 
instrument is to be returned to Hanna Instruments, 
first obtain a Returned Goods Authorization (RGA) 
number from the Technical Service department and 
then send it with shipping costs prepaid. When 
shipping any instrument, make sure it is properly 
packaged for complete protection.

• Pour 3” of standard 

HI7031

 into a beaker.

• Place the probe in calibration solution 

HI7031

The probe tip should be centered in the solution 
and submersed 2” (5 cm). The probe should be 
kept at least 1” from the walls and bottom of the 
beaker. The meter will automatically recognize 
the solution.

• If the solution is not recognized or is out of the 

accepted range “---- WRNG“ is displayed. 

• If the standard is recognized “REC“ is displayed 

until the reading is stable and the calibration is 
accepted.

• After acceptance, the “Stor” message is displayed 

and meter returns to measurement mode.

 

Clear calibration

Place meter in calibration mode. Press and hold 
ON/OFF until “Clr” is displayed. The meter will now 
be at default calibration. 
After restarting the meter the “Err CAL” message 
will appear (for 2 seconds) until a new calibration 
is performed.

 

Error Message

If the measured temperature is higher than 
50.0 ºC or lower than 0 ºC, the 50.0 ºC or 0.0 ºC 
temperature value will blink on the LCD in measure 
mode. In both cases, the “Err TEMP” message will 
appear. If CAL or HOLD mode was entered by the 
user the meter will enter in measurement mode. 

To reenter Setup press Setup button again.

To select EC unit

To select the measurement unit when “EC UNIT” 
is displayed press the ON/OFF briefly to change 
between mS/cm, dS/m or μS/cm.

To select the temperature unit (°C/°F)

To select the temperature unit when “TEMP” is 
displayed press the ON/OFF briefly to change 
between °C or °F.

To select the Auto-Off time

To select the AUTO OFF when “AOFF” is displayed 
press the ON/OFF briefly to change between 8 min, 
60 min or --- (disabled).

To select the HOLD mode

To select the Hold mode when “HOLD” is displayed 
press the ON/OFF briefly to change between 
disabled “no“, Hold mode “YES“ or Stability mode 
“StAb“.

To return to measurement mode

Press and hold the Setup button for 1 second to 
exit the menu (or press and hold ON/OFF button for 
2 seconds after the last parameter was selected).

HI7031-023

1413 μS/cm (1.41 mS/cm) 

GroLine Conductivity Standard 

Solution, 230 mL bottle

HI7031M

1413 μS/cm (1.41 mS/cm) 

calibration solution, 230 mL

  HI7031L 

 1413 μS/cm (1.41 mS/cm) 

 calibration solution, 500 mL

HI70031P

1413 μS/cm (1.41 mS/cm) 

calibration solution, 

20 mL sachets (25 pcs.)

Accessories

  IST98331   07/18

Setup 

Button

Battery Replacement

Replace the battery when percentage indicator 
approaches 0%. To change the CR2032 Li-ion 
battery, turn the battery cover located on the back 
of the meter counterclockwise to unlock. Remove 
cover and replace with new b side facing 
up. Push the ON/OFF button to start the meter.

Note: Batteries should only be replaced in a 

safe area using the battery type specified in 

this instruction manual. Old batteries should be 

disposed in accordance with local regulations.

For better accuracy, frequent calibration of the 
meter is recommended. In addition, the instrument 
must be recalibrated whenever:

• When high accuracy is required.

• At least once a month.

Measurement

Press the ON/OFF button to turn the meter ON. Place 
the probe to the desired depth and saturate the soil 
if needed with deionized water. The conductivity 
value, automatically compensated for temperature, 
will be shown on the LCD.
The meter displays conductivity in the selected unit.
The temperature measured by the probe will be 
shown on the secondary LCD.

Note: Before taking any measurement make 

sure the meter has been calibrated.

Calibration procedure

• Enter calibration mode from measurement 

mode. 

• The meter will enter the calibration mode, 

displaying “1.41 USE” .

Measurement and Calibration

Maintenance:

1) Rinse probe with water after use and between 

samples. Dry off.

2) Store GroLine with the protection cover over 

conductivity probe.

Содержание GroLine HI98331

Страница 1: ...INSTRUCTION MANUAL HI98331 Direct Soil Conductivity Temperature Meter with built in Stainless Steel Conductivity Probe ...

Страница 2: ...fice Each meter is supplied with CR2032 battery 1 pc Storage Protection sleeve Instruction manual Quality Certificate Note Save all packing material until you are sure that the instrument functions correctly All defective items must be returned in the original packaging with the supplied accessories General Description The Hanna Instruments GroLine is a special pocket conductivity meter expressly ...

Страница 3: ... button Operation Before using Hanna Instruments products make sure that they are entirely suitable for your specific application and for the environment in which they are used Avoid touching the probe at all times Any variation introduced by the user to the supplied equipment may degrade the instrument s performance Do not use or store the instrument in hazardous environments Recommendations for ...

Страница 4: ...model will be displayed Pressing the button 3 times will display the serial number SN 1 SN 2 SN 3 To enter measurement mode press the ON OFF button briefly Specifications Range 0 to 4000 µS cm 0 00 to 4 00 mS cm dS m 0 0 to 50 0 C 32 0 to 122 0 F Resolution 1 µS cm 0 01 mS cm dS m 0 1 C 0 1 F Accuracy 25 C 77 F 50 µS cm 0 to 2000 µS cm 300µS cm 2000to4000µS cm 0 05 mS cm 0 00 to 2 00 mS cm 0 30 mS...

Страница 5: ...return to measure mode The MEAS message appears and the meter returns to measure mode Taking another stable measurement is possible by pressing the ON OFF button briefly If the HOLD message appears then measured stable values EC and temp are already retained and hold on the display To enter calibration mode Press and hold down the ON OFF button until OFF is replaced by CAL Release the button To en...

Страница 6: ...ton for 1 second to exit the menu or press and hold ON OFF button for 2 seconds after the last parameter was selected For better accuracy frequent calibration of the meter is recommended In addition the instrument must be recalibrated whenever When high accuracy is required At least once a month Measurement PresstheON OFFbuttontoturnthemeterON Place the probe to the desired depth and saturate the ...

Страница 7: ...d and meter returns to measurement mode Clear calibration Place meter in calibration mode Press and hold ON OFF until Clr is displayed The meter will now be at default calibration After restarting the meter the Err CAL message will appear for 2 seconds until a new calibration is performed Error Message If the measured temperature is higher than 50 0 ºC or lower than 0 ºC the 50 0 ºC or 0 0 ºC temp...

Страница 8: ...partment and then send it with shipping costs prepaid When shipping any instrument make sure it is properly packaged for complete protection HI7031 023 1413 μS cm 1 41 mS cm GroLine Conductivity Standard Solution 230 mL bottle HI7031M 1413 μS cm 1 41 mS cm calibration solution 230 mL HI7031L 1413 μS cm 1 41 mS cm calibration solution 500 mL HI70031P 1413 μS cm 1 41 mS cm calibration solution 20 mL...

Отзывы:

Похожие инструкции для GroLine HI98331